Siemens Healthineers Biograph Vision

Workstation Freezes or Loses Communication

On this page

Asset Type

PET / CT System

Manufacturer

Siemens Healthineers

Model

Biograph Vision

What This Guide Helps With

Use this guide when the operator workstation becomes unresponsive or loses communication with scanner components because of external connections, network conditions, or software-state issues.

Step-by-Step Troubleshooting

1. Protect the Patient and Stop Dependence on an Unresponsive Workstation

If the workstation freezes during patient positioning or acquisition, do not assume the scanner is safely controllable.

Maintain patient safety, stop further examination activity, and use appropriate approved system controls to place the patient in a safe condition.

Expected outcome: The patient is safe and no active procedure relies on an unresponsive workstation.

2. Define the Failure

Determine whether:

Record any message displayed.

Expected outcome: The failure is distinguished as local workstation, application, or scanner-communication related.

3. Check Display and Input Devices

Inspect the workstation's accessible external components:

A blank display can be mistaken for a frozen computer.

Expected outcome: Display and input hardware are powered, connected, and functional.

If an external peripheral connection correction restores normal control, verify operation and stop.

4. Inspect Accessible Network and Communication Connections

Check accessible Ethernet and system communication connections for:

Do not move the scanner to another network port or VLAN without authorization.

Expected outcome: Accessible communication connections are intact and correctly connected.

If reseating an approved external connection restores stable communication, verify operation and stop.

5. Determine Whether the Problem Is Local or Infrastructure-Wide

Check whether other connected imaging or clinical systems are experiencing similar communication issues.

If multiple systems are affected, coordinate with the appropriate network or infrastructure team rather than altering the scanner configuration.

Expected outcome: A local scanner problem is distinguished from a broader infrastructure problem.

6. Allow an Application to Recover if Appropriate

If the workstation is temporarily busy rather than completely nonresponsive, avoid repeated clicking or forced power removal.

Observe whether normal operation returns within the expected local workflow.

Expected outcome: The application either recovers normally or remains clearly unresponsive.

If it recovers and communication remains stable during verification, troubleshooting may stop.

7. Perform a Controlled Restart if Required

If the system is not performing an active patient acquisition and approved normal controls permit it, perform one controlled workstation or system restart according to facility and manufacturer-approved practice.

Do not use hard power removal unless specifically required by an approved procedure.

Expected outcome: The workstation restarts normally and reconnects to required scanner subsystems.

If normal operation and communication are restored and remain stable, proceed to verification.

8. Verify Scanner Communication

Confirm after recovery that the workstation correctly communicates with required scanner components and can:

Expected outcome: Communication remains stable without repeat disconnects.

9. Perform Final Functional Verification

Verify:

Expected outcome: The workstation and scanner communicate reliably through required functions.

If all checks pass, troubleshooting is complete.

10. Escalate Recurring Freezes or Communication Loss

If the workstation repeatedly freezes, crashes, or loses scanner communication after connections and infrastructure conditions are verified, stop external troubleshooting.

Expected outcome: The affected system is removed from clinical use and escalated for qualified service.

If the Problem Persists

Common external causes have been ruled out. Remaining possibilities may involve workstation hardware, operating-system or application faults, internal scanner networking, system controllers, protected network configuration, data corruption, or other service-level conditions.

The device should be:

Do not modify protected network settings, reinstall system software, or enter unauthorized service utilities.

Required scanner and workstation functionality must be verified before return to clinical use.

Knowing when to stop external troubleshooting and escalate is proper troubleshooting.

Clinical Use Tip

A workstation that intermittently reconnects after freezing should not be considered reliable until communication remains stable through an appropriate functional test.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported that the Biograph Vision workstation lost communication with the scanner during setup."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an accessible Ethernet connection at the workstation partially seated."

Resolution

What action was taken.

Example:
"Clinical Engineering secured the connection, restarted the workstation normally, verified stable scanner communication and workflow operation, and returned the system to service."

Helpful Details to Include (If Known)

Final Thought

Protect the patient first, separate a simple peripheral or network connection issue from a true system fault, verify sustained communication after correction, and escalate recurrent workstation instability.

That is successful troubleshooting.