Siemens Healthineers Cios Spin

DICOM Send, PACS, or Worklist Failure

On this page

Asset Type

C-Arm

Manufacturer

Siemens Healthineers

Model

Cios Spin

What This Guide Helps With

Addresses failed image transfer or worklist retrieval caused by network connectivity, destination availability, workflow selection, configuration, or infrastructure problems.

Step-by-Step Troubleshooting

1. Protect Patient Care and Preserve Clinical Data

A DICOM or worklist problem usually does not require interruption of safe image acquisition, but do not assume images are archived merely because they are visible locally.

Ensure the clinical team understands the communication failure and follows the facility's approved process for retaining images and documenting patient information until connectivity is restored.

Do not delete unsent studies while troubleshooting.

Expected outcome: Clinical data are preserved and patient care can continue using the approved downtime workflow.

2. Confirm Which DICOM Function Is Failing

Determine whether the failure affects:

Record any displayed failure message exactly.

Expected outcome: The issue is narrowed to a specific DICOM function or destination.

3. Verify the Local Network Connection

Inspect the Cios Spin's accessible network connection.

Check:

Use a known-good approved network cable or known-good port comparison when appropriate and authorized.

Expected outcome: A valid physical network link is present. If restoring the connection restores DICOM service, proceed to end-to-end verification.

4. Confirm Other System Communication Is Normal

Verify the Cios Spin is otherwise operating normally and that workstation communication is stable.

If local workstation or system communication is also failing, address that broader system issue before focusing on PACS.

Expected outcome: The problem is isolated to network/DICOM communication rather than a general workstation failure.

5. Verify Patient and Workflow Selection

Confirm that the correct patient, exam, and destination workflow were selected.

For worklist problems, confirm search criteria or filters are not unintentionally excluding the expected patient.

Do not create duplicate patient records simply to bypass a worklist problem unless the facility's approved downtime workflow requires it.

Expected outcome: User-level workflow conditions are not blocking worklist retrieval or image sending.

6. Determine Whether the Problem Is Device-Specific

When possible, compare with another networked imaging device on the same network area or confirm with PACS/IT personnel whether the destination is available.

If multiple modalities cannot reach the same PACS or worklist service, the issue is likely infrastructure-side rather than specific to the Cios Spin.

Expected outcome: The failure is localized to the Cios Spin or identified as a broader network/PACS issue.

7. Review Operator-Accessible DICOM Status

Check the system's accessible send queue, study status, or communication indicators where available.

Determine whether studies are:

Do not alter DICOM AE titles, addresses, ports, or protected network configuration unless the change is authorized and based on verified configuration records.

Expected outcome: The communication state is understood without introducing configuration changes.

8. Coordinate Infrastructure Checks

If the physical network connection is valid, coordinate with PACS or network support to verify:

Clinical Engineering should provide the device identity and observed behavior without guessing at network configuration.

Expected outcome: Infrastructure-side problems are identified or ruled out.

9. Perform End-to-End DICOM Verification

After correction, perform the appropriate approved test:

Expected outcome: DICOM communication completes end to end. Troubleshooting can stop.

10. Escalate Unresolved Device-Side DICOM Problems

If the network path and destination are verified but the Cios Spin still cannot retrieve worklists or send images, escalate for service-level network or DICOM configuration evaluation.

Do not perform undocumented configuration changes or factory resets.

Expected outcome: The device is appropriately restricted or removed from service if required by facility workflow until reliable image transfer is restored.

If the Problem Persists

Once the physical network, workflow, destination availability, and infrastructure have been ruled out, remaining causes may involve DICOM configuration, software services, network-interface hardware, stored credentials or certificates where applicable, or service-level system configuration.

The Cios Spin should be:

Before normal use resumes, confirm both image transfer and worklist functionality as applicable and ensure required studies have reached their intended destination.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

A locally saved image is not the same as a successfully archived image; verify PACS receipt before considering the transfer problem resolved.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported the Cios Spin could acquire images but studies remained pending and would not transmit to PACS."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the external Ethernet cable had been connected to an inactive wall network port after the system was moved."

Resolution

What action was taken.

Example:
"The system was connected to the verified imaging-network port, pending images transmitted successfully, and receipt was confirmed at PACS."

Helpful Details to Include (If Known)

Final Thought

Preserve clinical data, verify the physical network and workflow before changing configuration, coordinate appropriately with IT or PACS teams, and confirm the complete DICOM path before closing the work order.

That is successful troubleshooting.