Siemens Healthineers Cios Spin

Poor 2D Image Quality, Excessive Noise, or Artifact

On this page

Asset Type

C-Arm

Manufacturer

Siemens Healthineers

Model

Cios Spin

What This Guide Helps With

Addresses degraded 2D images caused by positioning, obstruction, contamination, accessories, technique selection, detector conditions, motion, or external environmental factors.

Step-by-Step Troubleshooting

1. Protect the Patient and Avoid Unnecessary Exposure

Do not repeatedly expose a patient simply to troubleshoot poor image quality. If adequate imaging cannot be obtained without unnecessary repeat exposure, move the procedure to another verified imaging system.

Use a phantom or approved test object for technical evaluation whenever possible.

Expected outcome: Troubleshooting does not create unnecessary radiation exposure.

2. Define the Image-Quality Complaint

Determine whether the issue is:

Save or review representative images when permitted.

Expected outcome: The image defect is characterized well enough to test logically.

3. Inspect the Imaging Path

Verify that nothing unintended is in the X-ray beam.

Look for:

Expected outcome: The imaging field contains only intended anatomy or test material and required clinical equipment. If removal of an obstruction resolves the artifact, verify and stop.

4. Clean Accessible Imaging Surfaces

Inspect externally accessible detector covers and relevant surfaces for contamination, residue, adhesive, cleaning streaks, or debris.

Clean only according to approved facility and manufacturer-compatible practices.

Do not open the detector or remove protective assemblies.

Expected outcome: External contamination is not creating the visible artifact.

5. Verify Positioning and Geometry

Confirm the C-arm, detector, patient or phantom, and imaging target are positioned appropriately.

Check for:

Expected outcome: Image geometry is appropriate and repeatable.

6. Review Operator-Accessible Imaging Settings

Confirm that the selected clinical protocol, imaging mode, dose/image-quality selection, and other operator-accessible settings are appropriate for the intended exam.

Compare with a known-good workflow or equivalent system when possible.

Do not alter calibration data or protected service parameters.

Expected outcome: An inappropriate user-level setting is ruled out or corrected. If image quality returns to normal, continue to verification.

7. Check for Patient or Object Motion

If images are blurred or inconsistent, determine whether motion is occurring from:

For technical testing, use a stationary phantom to separate system performance from patient motion.

Expected outcome: Motion-related degradation is identified or ruled out.

8. Compare Images Across Positions or Modes

Acquire approved test images using a consistent phantom while varying only one external condition at a time.

Determine whether the artifact:

Expected outcome: The issue is narrowed to an external condition or shown to require service-level evaluation.

9. Perform Final Image-Quality Verification

After correcting an external cause, obtain repeat images using the same test object and comparable settings.

Verify:

Use approved quantitative image-quality testing when required by facility procedure.

Expected outcome: Image quality is consistently acceptable and the correction is verified. Troubleshooting can stop.

10. Escalate Persistent Image Defects

Remove the Cios Spin from service if unexplained artifacts, abnormal noise, nonuniformity, or other image-quality defects remain after external causes are ruled out.

Do not attempt detector calibration, generator calibration, internal adjustment, or protected service procedures without proper authorization and documentation.

Expected outcome: A system with potentially diagnostic image-quality degradation is not returned to clinical use.

If the Problem Persists

After imaging-path obstructions, contamination, positioning, motion, user-accessible settings, and accessories have been eliminated, remaining causes may involve detector calibration, X-ray generation, image processing, detector electronics, software, or service-level configuration.

The Cios Spin should be:

Complete applicable image-quality, radiation, and functional testing before clinical return.

Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Use a phantom rather than repeated patient exposures when determining whether poor image quality comes from the system or the clinical setup.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported a persistent dark artifact on Cios Spin 2D fluoroscopic images."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found adhesive residue on the exterior detector cover in the imaging field."

Resolution

What action was taken.

Example:
"The detector cover was cleaned using an approved method, repeat phantom images showed the artifact was gone, and image quality was verified before return to service."

Helpful Details to Include (If Known)

Final Thought

Avoid unnecessary patient exposure, eliminate positioning and imaging-path causes first, verify corrections with a controlled test object, and escalate persistent image defects for qualified evaluation.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide