Siemens Healthineers CLINITEK Advantus

Strip Transport Or Jam Error

On this page

Asset Type

Urine Analyzer

Manufacturer

Siemens Healthineers

Model

CLINITEK Advantus

What This Guide Helps With

Troubleshooting strip feed, transport, tray movement, or jam errors caused by strips, contamination, alignment, or mechanical obstruction.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the analyzer is not actively processing patient samples before troubleshooting.

If patient testing is urgent, have staff use an alternate analyzer or approved backup method.

Expected outcome: Troubleshooting is performed without delaying patient care or risking incorrect results.

Stop Testing and Preserve the Sample Workflow

Ask the operator whether any patient strips were in process when the jam occurred.

Confirm whether results were completed, aborted, or need to be repeated according to lab procedure.

Expected outcome: No questionable or incomplete patient result is released.

Verify the Exact Error and When It Occurs

Observe the displayed message or operator report.

Note whether the jam occurs:

Expected outcome: The failure pattern is clear before parts or settings are suspected.

Power Down Before Clearing Any Visible Jam

If a strip is visibly stuck, power the analyzer down before attempting removal.

Do not force the strip transport mechanism, tray, or carrier.

Expected outcome: The analyzer is safe to inspect and the transport mechanism is not damaged during clearing.

Check for a Misloaded, Bent, Wet, or Damaged Strip

Inspect the strip path for:

Remove only loose or easily accessible material.

Expected outcome: Any obvious strip obstruction is removed without forcing the mechanism.

If this resolves the issue, clean the area as appropriate, restart the analyzer, verify operation, document the finding, and stop.

Inspect the Strip Loading Area

Check the strip entry area and transport surface for dried urine, crystallization, debris, paper lint, glove powder, or reagent pad residue.

Clean accessible exterior and operator-cleanable areas according to facility procedure and manufacturer-approved cleaning guidance.

Expected outcome: The strip path is clean enough for smooth strip movement.

Why it matters: Dried urine and reagent residue can create drag, cause sensor detection problems, or prevent proper strip advancement.

Check the Strip Table, Carrier, or Transport Area for Obstruction

Inspect the visible strip transport area for:

Reseat only user-removable components that are intended to be removed for routine cleaning.

Expected outcome: The transport area is clear and properly seated.

Verify the Correct Reagent Strips Are Being Used

Confirm staff are using the appropriate Siemens CLINITEK-compatible urine reagent strips for the analyzer and facility workflow.

Check that strips are not expired, warped, exposed to excess humidity, or stored improperly.

Expected outcome: The analyzer is being fed with valid, flat, dry strips.

Why it matters: Warped or humidity-damaged strips may not travel correctly through the mechanism.

Check Operator Technique

Ask staff to demonstrate how the strip is prepared and inserted.

Look for:

Expected outcome: Strip insertion matches the analyzer’s expected workflow.

If correcting technique resolves the issue, document user education and stop.

Restart the Analyzer

After clearing visible debris and confirming proper setup, power the analyzer off, wait briefly, and power it back on.

Allow the startup/self-check process to complete.

Expected outcome: The analyzer initializes without a transport, tray, or jam error.

Run a Non-Patient Verification Check

Use an approved control or test procedure per laboratory policy.

Observe whether the strip advances normally through the transport path.

Expected outcome: The strip feeds, advances, reads, and exits without hesitation, skewing, or repeat jam errors.

Check for Repeated Error With Multiple Valid Strips

If the error repeats, try more than one valid strip from an acceptable container or lot, if available.

Expected outcome: This helps separate a bad strip condition from a recurring analyzer transport issue.

Listen and Observe During Transport

Watch and listen during strip movement.

Note:

Expected outcome: Any mechanical pattern is documented for repair evaluation.

Do Not Disassemble the Transport Mechanism Internally

Do not open internal covers, adjust sensors, lubricate mechanisms, bypass switches, or attempt motor-level repair unless authorized by service documentation and facility policy.

Expected outcome: Troubleshooting remains within safe Clinical Engineering limits.

Determine Device Status

If the analyzer passes verification testing and no further jams occur, return it to service per facility procedure.

If the jam error repeats after external checks, remove the analyzer from service.

Expected outcome: Only a reliable analyzer is returned for patient testing.

If the Problem Persists

If the strip transport or jam error continues after checking strips, loading technique, visible debris, removable parts, cleaning condition, and restart behavior, common external causes have been ruled out.

The issue may involve an internal transport motor, drive component, sensor, alignment problem, or contamination beyond operator-accessible areas.

The device should be:

Knowing when to stop is proper troubleshooting. Repeated strip jams can create invalid or incomplete patient results and should not be worked around.

Clinical Use Tip

Do not troubleshoot strip transport problems while patient testing is actively in progress. Have the lab move testing to another analyzer or backup process first. Any strip involved in a transport fault should be treated as unreliable unless the laboratory’s procedure clearly states otherwise.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Lab reported the Siemens CLINITEK Advantus displayed a strip transport jam error during urine strip processing."

Cause

What was observed during troubleshooting.

Example:
"Found dried residue and a partially misfed reagent strip in the accessible strip transport area."

Resolution

What action was taken.

Example:
"Removed the strip, cleaned the accessible transport area, restarted the analyzer, ran verification testing, and returned the unit to service after normal strip advancement was confirmed."

Helpful Details to Include (If Known)

Final Thought

Strip transport problems on a urine analyzer should be approached carefully because they can affect sample timing, strip positioning, and result reliability. Start with patient safety, strip condition, operator technique, visible contamination, and basic reset steps before suspecting internal failure. If the problem repeats after external causes are ruled out, remove the analyzer from service and document clearly using the CCR method.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide