Siemens Healthineers CLINITEK Status+

Test Strip Feed / Positioning Error

On this page

Asset Type

Urine Analyzer

Manufacturer

Siemens Healthineers

Model

CLINITEK Status+

What This Guide Helps With

Troubleshooting strip feed, strip placement, test table movement, or positioning errors that prevent proper CLINITEK Status+ urinalysis testing.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the analyzer is not actively being used for patient testing.

If testing is needed immediately, have staff use another available urine analyzer or approved backup testing process.

Expected outcome: Troubleshooting is performed without delaying patient care or risking inaccurate results.

Verify the Reported Issue

Ask staff what occurred during testing:

Expected outcome: The problem is confirmed as a strip handling or positioning issue, not a general startup, display, or chemistry result issue.

Check for Incorrect Strip Placement

Verify the test strip was placed flat and correctly on the test table insert.

Confirm the strip was not angled, curled, upside down, reversed, or pushed outside the guide area.

Expected outcome: If incorrect placement caused the issue, reload a new strip correctly and repeat the test. If the analyzer accepts the strip, stop.

Confirm the Correct Strip Type Is Being Used

Verify staff are using Siemens-compatible strips approved for the CLINITEK Status+.

Also check that the analyzer setup matches the strip type being used.

Expected outcome: If the wrong strip type or setup selection was used, correct it, repeat with a valid strip, and stop if testing proceeds normally.

Inspect the Test Strip Condition

Check whether the strip is expired, warped, wet before use, contaminated, discolored, or left exposed to room humidity.

Siemens notes that the CLINITEK Status+ performs automatic checks on Siemens test strips, including strip identification and humidity exposure checks.

Expected outcome: If questionable strips are found, discard them per facility policy, test with a fresh properly stored strip, and stop if the error clears.

Inspect the Test Table and Insert

Remove the test table slowly and inspect the table insert for:

Expected outcome: The test table and insert should be clean, seated correctly, and able to move without obstruction.

Clean the Test Table and Insert

Remove the test table and table insert. Clean the test table and insert using water and a cotton-tipped swab or facility-approved method, avoiding the white calibration bar. Siemens cleaning instructions identify removing the table, lifting out the insert, draining the drip tray if needed, and cleaning the test table and insert while avoiding the white calibration bar.

Dry the parts thoroughly before reinstalling.

Expected outcome: If residue or contamination was causing strip drag or misalignment, the device should resume normal strip positioning after cleaning.

Inspect the White Calibration Bar

Visually inspect the white calibration bar for obvious marks, scratches, urine residue, discoloration, or damage.

Do not scrub or clean the calibration bar unless directed by the manufacturer’s procedure. Siemens instructions specifically caution cleaning the table and insert except for the white calibration bar.

Expected outcome: If the calibration bar is visibly damaged or contaminated, remove the analyzer from service for evaluation.

Reinstall the Test Table Correctly

Reinsert the test table with the white calibration bar facing up. The operator guidance states to insert the test table while holding the end opposite the white calibration bar and to push it into the analyzer just over halfway, not fully into the analyzer.

Expected outcome: The table should seat properly and move normally. If the table was inserted too far or incorrectly, reinstalling it correctly may resolve the feed or positioning error.

Check for Obstruction in the Table Path

With the test table removed, inspect the accessible table opening for visible debris, broken strip material, dried sample, paper, labels, or foreign objects.

Do not force tools deep into the analyzer or disassemble internal components.

Expected outcome: If a visible obstruction is safely removed and the table moves normally afterward, repeat a test with a fresh strip and stop if successful.

Power Cycle the Analyzer

Turn the analyzer off, wait briefly, then restart it.

Observe whether the test table initializes normally and whether the positioning error returns.

Expected outcome: If the issue was a temporary movement or detection state, the analyzer may recover after restart.

Repeat With a Known-Good Strip and Clean Table

Use a fresh, properly stored strip and observe the full test sequence.

Watch for abnormal movement, hesitation, grinding, binding, repeated table movement, or failure to detect the strip.

Expected outcome: A successful test confirms the problem was likely caused by strip placement, strip condition, table contamination, or setup issue.

Check for Repeated Mechanical Symptoms

If the table does not move, moves unevenly, jams, clicks, binds, or repeatedly fails with clean parts and correct strips, suspect a test table drive, sensor, or internal positioning fault.

Expected outcome: The analyzer should be removed from service instead of repeatedly forcing tests.

If the Problem Persists

If the CLINITEK Status+ continues to show test strip feed or positioning errors after correct strip placement, fresh strips, cleaning, proper table installation, and basic restart checks, common external causes have been ruled out.

Remove the analyzer from service, label it Out of Service, and send it for repair or bench evaluation.

Do not continue patient testing on a unit that cannot reliably position or read strips. Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Do not troubleshoot strip feed or positioning issues during active patient testing. Move testing to another analyzer or approved backup method first. A mispositioned strip can lead to failed testing, delayed results, or questionable accuracy.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Clinical staff reported the CLINITEK Status+ would not properly feed or position urine test strips during patient testing."

Cause

What was observed during troubleshooting.

Example:
"Test table insert had dried urine residue and debris causing the strip to drag and misalign during the test sequence."

Resolution

What action was taken.

Example:
"Removed analyzer from patient use, cleaned and dried the test table and insert, reinstalled the table correctly, repeated testing with a fresh strip, and verified normal strip positioning."

Helpful Details to Include (If Known)

Final Thought

Strip feed and positioning problems should be handled logically. Start with patient safety, then verify strip placement, strip condition, test table cleanliness, and correct table installation. If the analyzer still cannot position strips reliably, escalation is appropriate because accurate urinalysis depends on consistent strip timing and alignment. Clear CCR documentation supports both repair follow-up and patient safety.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide