Siemens Healthineers Multix Impact

Workstation, DICOM Send, Worklist, or Archive Communication Failure

On this page

Asset Type

Radiographic System

Manufacturer

Siemens Healthineers

Model

Multix Impact

What This Guide Helps With

Troubleshooting Multix Impact workstation, DICOM send, worklist, archive, network, or image transfer failures before repair escalation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the Multix Impact is not actively acquiring patient images before restarting the workstation, disconnecting cables, or changing network connections.

If images were already acquired, verify with Radiology staff whether the images are stored locally and whether patient care depends on immediate image availability.

Expected outcome: Patient imaging and completed studies are protected before troubleshooting begins.

Confirm the Exact Communication Failure

Ask the technologist what failed and when it occurred.

Determine whether the issue involves:

Expected outcome: The issue is separated into worklist, image send, archive, workstation, or network-related failure.

Check Whether the System Can Still Acquire Images

Confirm whether the system can still perform exposures and store images locally.

Do not assume the entire radiographic system has failed if only DICOM, worklist, or archive communication is affected.

Expected outcome: The issue is identified as either a communication problem only or a broader workstation/system problem.

Verify the Workstation Is Powered On and Responsive

Check that the acquisition workstation, monitor, keyboard, mouse, touch interface, and system console are powered and responding normally.

If the workstation is frozen, coordinate with the technologist before restarting. Confirm no active acquisition or unsaved image review is in progress.

Expected outcome: The workstation is operational enough to review studies, queues, and network status.

Check for Obvious Application Messages

Review the workstation screen for DICOM, network, archive, worklist, database, or storage-related alerts.

Note the exact wording of any error message before clearing or restarting anything.

Expected outcome: The reported issue is matched to a visible system message that can be documented or escalated accurately.

Verify Network Cable Connection if Wired

Inspect the network cable at the workstation, wall jack, floor box, or network drop if accessible.

Confirm the cable is fully seated, not damaged, not stretched, and connected to the correct network port.

If available, check for link/activity lights at the network port.

Expected outcome: A loose or disconnected network cable is corrected. If communication returns, document the cable issue and stop.

Check Wireless Network Status if Applicable

If the Multix Impact workstation uses wireless communication for worklist, PACS, or archive access, check whether the wireless connection shows connected status.

Confirm the issue is not isolated to a known wireless dead zone, access point issue, or temporary network outage.

Expected outcome: Wireless status is confirmed before assuming workstation or DICOM configuration failure.

Compare With Other Imaging Systems

Ask Radiology staff whether other X-ray rooms, mobile systems, CT, or other modalities are also having worklist or PACS send issues.

If multiple systems are affected, the problem may be network, PACS, RIS, worklist server, or hospital interface related rather than a Multix Impact failure.

Expected outcome: The issue is narrowed to this system only or identified as a broader department/network problem.

Check Worklist Function Separately From DICOM Send

Attempt to refresh or query the worklist if site workflow allows.

If worklist fails but DICOM send still works, the problem may involve RIS, modality worklist, network routing, or worklist configuration.

If DICOM send fails but worklist works, the problem may involve PACS/archive destination, DICOM queue, storage commitment, or destination availability.

Expected outcome: Worklist and image transfer are evaluated as separate communication paths.

Verify Patient and Exam Selection

Confirm that the exam was selected correctly and that patient demographics are complete enough for the expected workflow.

Incorrect, incomplete, or manually entered demographics can sometimes cause study matching, archive, or PACS routing problems.

Expected outcome: The failure is not due to incorrect patient selection, incomplete exam information, or workflow mismatch.

Review the DICOM Send Queue

Check whether images are queued, failed, pending, or already marked as sent.

If images are stuck in queue, note whether all destinations are failing or only one PACS/archive destination is affected.

Expected outcome: The problem is identified as a queue issue, destination issue, or failed send attempt.

Retry Send Only When Appropriate

If the workstation allows a normal resend and Radiology workflow permits it, retry sending one affected study or image set.

Avoid repeated blind resends if the queue is backing up, because this may create duplicate studies or overload the destination.

Expected outcome: A temporary send failure clears, or the repeated failure is confirmed for documentation and escalation.

Check Local Storage or Database Warnings

Look for messages indicating local storage full, database error, archive unavailable, or exam save failure.

A workstation with limited local storage or database problems may still acquire images but fail to process, send, or archive normally.

Expected outcome: Storage or database warnings are identified before deeper troubleshooting.

Restart the Workstation Only After Protecting Patient Data

If the workstation is unresponsive or communication services appear locked, restart only after confirming with Radiology staff that no active exam, unsaved image, or urgent patient workflow will be interrupted.

After restart, verify application startup, worklist query, and DICOM send status.

Expected outcome: A software or communication service lockup clears. If communication returns, document the restart and successful verification.

Check Date, Time, and System Identification

Confirm the workstation date and time appear correct.

Incorrect system time or identification can interfere with worklist queries, study matching, archive routing, or network authentication.

Expected outcome: Obvious date/time mismatch is ruled out or documented for escalation.

Verify Destination-Specific Failure

If multiple DICOM destinations are configured, check whether failure occurs with PACS only, archive only, printer only, or all destinations.

A single failed destination usually points toward PACS/archive/network configuration or destination availability. All destinations failing may suggest workstation network communication failure.

Expected outcome: The failure is narrowed to one destination or the entire communication path.

Check for Recent Changes

Ask whether there were recent changes to the workstation, network jack, IP address, PACS, RIS, firewall, wireless access points, or hospital network.

Recent IT or PACS changes are common causes of sudden worklist or DICOM communication failure.

Expected outcome: A change-related cause is identified or ruled out.

Coordinate With PACS, IT, or Radiology Informatics

If external checks are normal but communication still fails, contact the appropriate support group with the system name, room, IP address if available, destination affected, error message, and time of failure.

Expected outcome: Network, PACS, RIS, or DICOM destination issues are escalated with useful details.

Verify Function After Correction

After any correction, confirm the workstation can perform the affected function.

Depending on the original complaint, verify:

Expected outcome: The reported communication function is restored and confirmed with Radiology staff.

If the Problem Persists

If the Multix Impact still has workstation, DICOM, worklist, or archive communication failure after checking power, workstation responsiveness, network connection, wireless status, queue behavior, destination status, and broader network/PACS availability, common external causes have been ruled out.

The issue may involve workstation software, DICOM configuration, network adapter failure, local database problems, application services, PACS/RIS interface issues, or another internal/system-level fault.

The device should be:

Knowing when to stop is proper troubleshooting. Repeated restarts, repeated resends, or unapproved configuration changes can create duplicate studies, lost images, or delayed care.

Clinical Use Tip

Do not troubleshoot communication, restart the workstation, or disconnect network cables during an active patient exam unless patient safety requires immediate action. If imaging must continue, have Radiology staff move the patient to another available system and protect any completed images before deeper troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Radiology reported the Siemens Multix Impact worklist would not load and completed exams were not sending to PACS."

Cause

What was observed during troubleshooting.

Example:
"Workstation was responsive, image acquisition was available, but DICOM send queue showed repeated failed transfers while the network connection and cable were verified."

Resolution

What action was taken.

Example:
"Verified no active patient exam, checked workstation and network connections, confirmed issue was isolated to PACS communication, escalated to Radiology Informatics/vendor support, and labeled system limited use pending repair."

Helpful Details to Include (If Known)

Final Thought

Multix Impact communication troubleshooting should follow a clear path: protect patient images first, confirm the exact failure, check workstation and network basics, separate worklist from DICOM send issues, and escalate when the problem moves beyond safe external checks. Good documentation helps Clinical Engineering, Radiology, PACS, IT, and vendor support resolve the issue faster.

That is successful troubleshooting.