Siemens Healthineers Symbia Intevo

Acquisition Workstation, DICOM, Archive, or RIS/PACS Communication Failure

On this page

Asset Type

SPECT / CT System

Manufacturer

Siemens Healthineers

Model

Symbia Intevo

What This Guide Helps With

Troubleshooting Symbia Intevo workstation, DICOM, archive, RIS/PACS, network, or study transfer failures using safe external checks first.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the Symbia Intevo is not actively scanning, reconstructing, or transferring a clinically urgent study.

If a patient is on the table or the exam is in progress, do not interrupt acquisition or workstation processes unless directed by qualified imaging staff.

Expected outcome: Troubleshooting does not interrupt imaging, delay care, or risk study data loss.

Confirm the Exact Communication Complaint

Ask Nuclear Medicine or imaging staff what failed.

Identify whether the issue involves:

Expected outcome: The issue is separated into workstation, local network, DICOM destination, RIS/worklist, PACS/archive, or workflow-related failure.

Verify the Study Is Complete and Available Locally

Have qualified imaging staff confirm the study exists on the acquisition workstation and can be opened locally.

Confirm whether raw data, reconstructed images, CT images, fused SPECT/CT images, or secondary captures are affected.

Expected outcome: The issue is confirmed as a communication/export problem, not a failed acquisition or missing local study.

If the study is not available locally, stop and escalate as an acquisition or workstation data issue.

Check for Active Patient or Study Processing

Confirm the workstation is not still reconstructing, processing, or finalizing the study.

Some transfers may not begin until the exam or reconstruction step is complete.

Expected outcome: A delayed transfer is not mistaken for a failed transfer.

If the transfer completes after processing finishes, document the delay and stop.

Check Workstation Power and Basic Operation

Confirm the acquisition workstation, monitors, keyboard, mouse, and any related computer peripherals are powered on and responsive.

Look for frozen screens, login prompts, application errors, or abnormal system behavior.

Expected outcome: The workstation is available for normal use before network or PACS troubleshooting begins.

Check Physical Network Connections

Inspect the Ethernet connection at the workstation, wall jack, network switch connection point if accessible, and any nearby patch cable.

Confirm the cable is fully seated and not damaged, stretched, crushed, or disconnected.

Expected outcome: A loose or damaged network connection is ruled out.

If reseating the cable restores communication, confirm successful DICOM or worklist function and stop.

Confirm Network Link Indicators

Check for link/activity lights at the workstation network port or connected network equipment if visible.

A missing link light may indicate a disconnected cable, bad patch cable, disabled wall jack, or network port issue.

Expected outcome: The device has a basic physical network link.

If there is no link light after cable and jack checks, escalate to IT/network support and document the finding.

Try a Known-Good Network Cable if Accessible

Replace the external Ethernet patch cable with a known-good cable if the workstation uses an accessible patch cable.

Do not move internal scanner, gantry, table, or cabinet wiring.

Expected outcome: A simple external cable failure is ruled out.

If the known-good cable restores communication, document the cable replacement and stop.

Confirm Other Networked Systems Are Working

Ask imaging staff whether other modalities in the department can send to PACS or receive RIS worklists.

If multiple devices are affected, the issue is likely network, RIS, PACS, archive, or enterprise-system related rather than isolated to the Symbia Intevo.

Expected outcome: The failure is classified as device-specific or broader system-wide.

Check Whether the Failure Is One Destination or All Destinations

Determine whether the Symbia Intevo can send to any DICOM destination.

Examples:

Expected outcome: The affected communication path is narrowed.

If only one destination fails, involve PACS/RIS/IT support for that destination while Clinical Engineering verifies the device-side basics.

Review Error Messages and Queue Status

Have imaging staff show the exact workstation message or DICOM queue status.

Look for errors such as failed association, destination unreachable, connection timeout, rejected study, invalid accession, storage commitment issue, or queue retry failure.

Expected outcome: The error supports either network failure, destination failure, configuration mismatch, or patient/study data mismatch.

Check for Study Demographic or Accession Problems

Verify with imaging staff that patient name, MRN, accession number, exam type, and study date match the expected order.

RIS/PACS failures can occur when the study information does not match the order or destination rules.

Expected outcome: The failure is not caused by incorrect demographics or mismatched order data.

If corrected demographics or order reconciliation allows transfer, confirm the study arrives in PACS and stop.

Confirm Modality Worklist Function

If RIS worklist is the complaint, confirm whether the workstation can query the worklist and whether the expected patient/order appears.

Check whether the issue affects all patients or only one order.

Expected outcome: A single missing order is separated from a full RIS/worklist communication failure.

If only one order is missing, escalate to Radiology IT, scheduling, or RIS support rather than replacing device hardware.

Check Date and Time Consistency

Confirm the workstation date and time appear reasonable and match hospital time closely.

Significant time mismatch can affect authentication, study matching, logs, and communication workflows.

Expected outcome: Communication failure is not being caused by an obvious time/date issue.

If the time is incorrect, follow approved site procedure for correction and retest communication.

Check for Recent Changes

Ask whether anything changed recently, including:

Expected outcome: Recent environmental or configuration changes are identified before suspecting scanner hardware.

Restart Only When Safe and Approved

If the workstation appears frozen or communication services appear stuck, coordinate with imaging staff before restarting any workstation application or system.

Confirm no acquisition, reconstruction, transfer, or patient exam is active.

Expected outcome: A safe restart clears a temporary workstation or application communication issue without risking data loss.

If communication works after restart, verify successful send/worklist/archive operation and document the restart.

Retest the Failed Workflow

Retest the exact failed function with imaging staff.

Examples:

Expected outcome: The original complaint is either resolved or reproducible.

Escalate Based on Failure Pattern

If physical network, workstation responsiveness, study completion, demographics, and basic workflow checks are acceptable, escalate appropriately.

Use this logic:

Expected outcome: The issue is routed to the correct support path without unnecessary hardware replacement.

If the Problem Persists

If the problem persists after external network checks, workstation status checks, study verification, queue review, demographic review, and safe restart, common external causes have been ruled out.

The issue may involve workstation software, DICOM configuration, network interface failure, RIS/PACS configuration, archive server communication, or Siemens application-level service problems.

The Symbia Intevo should be:

Knowing when to stop is proper troubleshooting. Clinical Engineering should not perform unauthorized workstation software changes, DICOM configuration edits, or internal system repair beyond approved support procedures.

Clinical Use Tip

Do not troubleshoot communication failures during an active scan unless patient safety requires immediate action. If studies cannot be reliably transferred, move the patient to a backup imaging workflow or alternate approved device path first, and have imaging staff follow the approved downtime or alternate-transfer workflow before continuing patient imaging. Therapy and imaging continuity should be maintained before troubleshooting continues.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nuclear Medicine reported the Siemens Symbia Intevo acquisition workstation would not send completed SPECT/CT studies to PACS."

Cause

What was observed during troubleshooting.

Example:
"Verified workstation was responsive and study was available locally, but DICOM send queue showed repeated connection failures to the PACS destination while other networked systems were also affected."

Resolution

What action was taken.

Example:
"Confirmed external network cabling was seated, documented PACS communication failure, removed device from routine clinical workflow pending resolution, and escalated to Radiology IT/PACS support and Siemens service."

Helpful Details to Include (If Known)

Final Thought

SPECT/CT communication troubleshooting should stay logical and safety-focused. Confirm the patient is safe, verify the study exists locally, check the workstation and network basics, then separate device-side issues from RIS, PACS, archive, or IT problems. Clear CCR documentation helps protect patient care and speeds escalation to the correct support team.

That is successful troubleshooting.