On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting Aera host startup, syngo software, reconstruction, or boot-related faults before escalation.
Step-by-Step Troubleshooting
Ensure Patient Safety First
Confirm no patient is inside the bore and no exam is actively in progress before troubleshooting host, console, or reconstruction issues.
Expected outcome: The MRI system is safe to evaluate without interrupting an active scan or delaying patient removal.
If a patient is involved, stop troubleshooting and coordinate safe patient removal through MRI staff.
Confirm the Exact Reported Problem
Ask MRI staff what failed and when it happened:
- Host will not boot
- syngo interface frozen
- Reconstruction delayed or failed
- Exam will not load
- System stuck during startup
- Scan completed but images are missing
- Error appeared after shutdown, restart, power event, or software freeze
Expected outcome: The fault is narrowed to host startup, application behavior, reconstruction, image transfer, or full system availability.
Check Console Power and External Power Conditions
Verify the host computer, monitors, keyboard, mouse, and console peripherals have power.
Check for:
- Monitor power lights
- Host power indicator
- UPS status, if present
- Tripped power strip or switched outlet
- Loose power cable behind the console
Expected outcome: Console hardware is powered normally.
If the issue is resolved by restoring external power or reconnecting a loose cable, verify normal startup and stop.
Check for Monitor or Display-Only Issues
If the system appears down, confirm whether the host is actually running but the display is blank.
Check:
- Monitor input/source selection
- Video cable connection
- Display power
- Multiple monitors, if equipped
- Whether keyboard lights respond
Expected outcome: A blank screen is not mistaken for a failed host.
If the host is running and only the display connection was loose or misconfigured, correct it, verify syngo display, and stop.
Check Keyboard, Mouse, and Console Input Devices
Confirm the keyboard and mouse are connected and responsive.
Look for:
- Loose USB connections
- Damaged cables
- Nonresponsive mouse or keyboard
- Accidental disconnection during cleaning or console movement
Expected outcome: The operator can interact with the host and syngo environment.
If input devices were disconnected or faulty, reconnect or replace externally, verify normal operation, and stop.
Verify Whether the Issue Is a Frozen Application or Full Host Lockup
Determine whether only syngo is frozen or the entire computer is nonresponsive.
Check whether:
- Mouse pointer moves
- Keyboard shortcuts respond
- Clock updates
- Other windows respond
- Error messages can be acknowledged
Expected outcome: Clinical Engineering can distinguish software freeze from full host failure.
If the host is partially responsive, coordinate with MRI staff and follow site-approved restart procedures. Do not force repeated restarts without documenting the behavior.
Confirm Reconstruction Queue or Image Availability
If scans completed but images are delayed or missing, ask staff whether the issue affects:
- One patient only
- One protocol or sequence
- All studies
- Only recon images
- Image transfer to PACS
- Local viewing on the scanner
Expected outcome: The issue is separated into reconstruction delay, local image issue, or network/PACS transfer issue.
If images are present locally but not in PACS, troubleshoot network/export workflow separately rather than assuming reconstruction failure.
Check Network and External Communication Indicators
For host or image transfer complaints, verify external network conditions before escalating.
Check:
- Network cable seated at console connection point
- Link lights at accessible network jack or switch port, if available
- Whether other modality network functions are affected
- Whether PACS, RIS, or worklist access is also failing
Expected outcome: A network interruption is ruled in or ruled out as the cause of missing worklist, PACS transfer, or archive behavior.
If a loose network cable or known network outage is found, document and route appropriately.
Ask About Recent Changes
Determine whether the fault started after:
- Power outage
- Emergency power testing
- Planned shutdown
- Software update
- Network maintenance
- HVAC or cooling issue
- Service visit
- Unexpected system restart
Expected outcome: The fault is tied to a possible external event or recent service condition.
This helps avoid treating a site power or network event as a random host failure.
Check Room and Equipment Environmental Conditions
Verify the scanner equipment area is not showing obvious environmental problems.
Check for:
- High room temperature
- Unusual heat at console or equipment cabinets
- Burning smell
- Abnormal fan noise
- Water leak near equipment
- Alarm indicators on nearby support equipment
Expected outcome: Obvious environmental or safety concerns are identified before further restart attempts.
If heat, smoke smell, water, or abnormal equipment behavior is present, stop and escalate immediately.
Perform One Controlled Restart Only If Appropriate
If the system is not actively scanning, no patient is involved, and site procedure allows it, coordinate with MRI staff before performing a controlled restart.
Expected outcome: The host, syngo environment, and reconstruction function return normally after a proper restart.
If the system boots normally and MRI staff confirm scan/recon operation, document the restart and final status.
Do not perform repeated power cycles if the same fault returns.
Verify System Readiness After Recovery
After any successful restart or correction, have MRI staff confirm:
- syngo opens normally
- Patient browser is accessible
- Protocols load
- Reconstruction completes
- Images display locally
- Worklist/PACS function is normal, if applicable
- No active system error remains
Expected outcome: The device is returned to service only after functional readiness is confirmed.
If the issue returns during verification, remove the system from service and escalate.
If the Problem Persists
If power, display, peripherals, network connections, environmental conditions, and a controlled restart have been checked, the remaining cause is likely internal host hardware, syngo software, reconstruction subsystem, storage, database, or system communication related.
The MRI system should be:
- Removed from clinical use
- Labeled Out of Service
- Escalated to Siemens service or qualified MRI service support
- Held for bench/service evaluation before patient scanning resumes
Knowing when to stop is proper troubleshooting. Repeated restarts or unsupported software repair attempts can increase downtime, risk data loss, or mask a larger system issue.
Clinical Use Tip
Do not troubleshoot host, reconstruction, or boot faults during an active patient exam unless the immediate goal is safe patient removal. If scanning capability or image reconstruction is uncertain, move the patient schedule to another available MRI system or delay scanning until the Aera is verified safe and functional.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"MRI staff reported the Siemens MAGNETOM Aera syngo host would not complete startup and reconstruction was unavailable for scheduled exams."
Cause
What was observed during troubleshooting.
Example:
"Verified console power, monitor connections, keyboard/mouse, network connection, and environmental conditions; fault persisted after one controlled restart."
Resolution
What action was taken.
Example:
"Removed MRI system from service, labeled Out of Service, documented startup behavior, and escalated to Siemens service for host/reconstruction system evaluation."
Helpful Details to Include (If Known)
- Whether a patient was involved
- Exact error message or startup screen observed
- Whether syngo froze or the entire host locked up
- Whether local images were available
- Whether PACS, worklist, or network functions were affected
- Host power and monitor status
- Keyboard and mouse response
- Network cable/link light status, if checked
- Any recent power outage, shutdown, update, or service visit
- Unusual heat, smell, noise, or room temperature issue
- Whether one controlled restart was attempted
- Final device status
Final Thought
Host, syngo, reconstruction, and boot faults can quickly affect scanner availability and patient throughput. Clinical Engineering should verify the simple external causes first, protect patient safety, avoid repeated unsupported resets, and escalate when the issue points to internal system or software failure. Clear CCR documentation helps service teams understand exactly what was checked before handoff.
That is successful troubleshooting.