Siemens Healthineers MAGNETOM Aera

Image Artifact, Ghosting, Shading, or Failed QA Scan

On this page

Asset Type

MRI System

Manufacturer

Siemens Healthineers

Model

MAGNETOM Aera

What This Guide Helps With

Troubleshooting image artifacts, ghosting, shading, distortion, RF noise, or failed MRI QA scans before escalating for service.

Step-by-Step Troubleshooting

Ensure Patient Safety First

If artifacts or image quality problems occur during a patient exam, do not continue scanning if diagnostic quality is questionable.

Move the patient to another MRI system if clinically necessary and available.

Expected outcome: Patient care is not dependent on images from a system with unresolved quality concerns.

Confirm the Artifact or QA Failure

Review the reported issue with MRI staff and determine whether the problem occurred during:

Expected outcome: The issue is narrowed to a repeatable symptom instead of a vague image quality complaint.

Check for External RF Interference

Inspect the MRI room for unauthorized or unusual items, including phones, smart watches, tablets, pumps, monitors, cables, tools, carts, or non-MR equipment.

Remove any questionable item from the scan room before repeating QA.

Expected outcome: If RF noise or artifact was caused by external interference, image quality improves after the item is removed.

If this resolves the issue, document the finding and stop.

Verify MRI Room Door and RF Shielding Conditions

Confirm the scanner room door fully closes and seals properly. Look for damaged door fingers, loose shielding contact points, or objects preventing full closure.

Expected outcome: The RF-shielded room is properly closed during scanning.

If poor door closure caused the artifact, correct the obstruction or report the door/shielding issue for repair.

Check Coil Selection and Coil Connection

Confirm the correct coil is selected for the exam or QA scan. Inspect the coil connector, table connector area, and coil cable routing for obvious damage, looseness, or contamination.

Reseat the coil connection if safe and appropriate.

Expected outcome: The coil is properly recognized and connected before scanning.

If the artifact follows one coil, remove that coil from service and test with another known-good coil.

Repeat the Scan With a Known-Good Coil

If the issue occurred with a patient coil, repeat a basic test or QA sequence using a known-good coil approved for the system.

Expected outcome: If the known-good coil produces normal images, the original coil or coil cable is suspect.

If this resolves the issue, remove the faulty coil from service and document the coil-specific failure.

Inspect the Phantom and QA Setup

For failed QA scans, verify the correct phantom is being used, positioned correctly, and filled according to site practice.

Check for phantom bubbles, leaks, incorrect orientation, poor centering, or movement during scanning.

Expected outcome: QA is repeated with a correctly positioned and stable phantom.

If the QA scan passes after correcting phantom setup, document the setup issue and stop.

Verify Table Position and Patient/Phantom Centering

Confirm the patient or phantom is properly centered at isocenter and positioned according to the intended protocol.

Poor centering can contribute to shading, signal drop-off, distortion, or failed uniformity results.

Expected outcome: The scan is repeated with proper positioning and improved image quality.

Check Protocol and Parameter Selection

Confirm MRI staff used the correct protocol, coil combination, field of view, slice position, acceleration settings, and scan orientation.

Compare the failed scan against a recently successful protocol if available.

Expected outcome: The issue is not caused by an incorrect or modified protocol.

If a protocol change caused the artifact, restore the correct protocol or involve the MRI lead/physicist as appropriate.

Look for Patient Motion or Physiologic Ghosting

For patient images, check whether ghosting aligns with phase direction and whether motion, breathing, swallowing, cardiac motion, or patient discomfort may have contributed.

Expected outcome: Motion-related artifacts are identified as exam-specific rather than system failure.

If motion caused the issue, document it as clinical scan behavior, not equipment failure.

Check for Metal or External Objects Near the Anatomy

Confirm there were no removable metal objects, clothing components, patches, ECG leads, jewelry, hair accessories, or external devices near the scan area.

Expected outcome: Artifact from external objects is ruled out before suspecting system hardware.

If removing the item resolves the artifact, document the cause and stop.

Review Recent Changes Around the Scanner

Ask whether any of the following changed recently:

Expected outcome: The artifact is correlated with a recent environmental, facility, or system change.

Check for System Warnings or Related Error Messages

Review the operator console for active warnings, coil messages, gradient warnings, cooling alerts, reconstruction errors, or QA-related messages.

Expected outcome: Any associated system warning is captured before escalation.

Do not clear or ignore recurring errors without documenting them.

Repeat Daily QA Under Controlled Conditions

If no obvious external issue is found, repeat the required QA scan using the correct phantom, proper setup, correct coil, and standard site procedure.

Expected outcome: QA either passes, confirming the issue was setup-related, or fails again, confirming a repeatable problem.

If QA fails repeatedly under controlled conditions, remove the system from clinical use according to site policy.

Compare Artifact Behavior Across Coils and Sequences

Determine whether the problem occurs:

Expected outcome: Clinical Engineering can provide Siemens service or the physicist with a clear failure pattern.

Escalate When Image Quality Cannot Be Verified

If image artifacts persist, QA fails repeatedly, or diagnostic image quality is uncertain, stop troubleshooting at the external-check level.

Expected outcome: The system is removed from clinical use until qualified service evaluation, physicist review, or manufacturer support confirms safe operation.

If the Problem Persists

If the MAGNETOM Aera continues to show artifacts, ghosting, shading, distortion, RF noise, or repeated QA failure after external causes have been ruled out, the issue may involve internal RF chain components, gradient performance, receive electronics, system calibration, shielding integrity, reconstruction, or environmental interference.

The MRI system should be:

Knowing when to stop is proper troubleshooting. MRI image quality problems can affect diagnosis and should not be treated as routine nuisance errors.

Clinical Use Tip

Do not troubleshoot image quality issues during an active patient exam unless patient safety requires immediate action. End the scan safely, move the patient to another available scanner if needed, and perform troubleshooting only when the magnet room and patient workflow are controlled.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"MRI staff reported ghosting and shading on MAGNETOM Aera images, with the daily phantom QA scan failing uniformity."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the QA phantom was not centered correctly and a non-MR accessory cart had been left near the scan room doorway."

Resolution

What action was taken.

Example:
"Removed the accessory cart, repositioned the phantom at isocenter, repeated QA successfully, and returned the MRI system to service."

Helpful Details to Include (If Known)

Final Thought

MRI artifact troubleshooting should move from patient safety to simple external checks before assuming a deeper system fault. Coils, positioning, RF interference, protocol selection, phantom setup, and room conditions should be ruled out first. If QA continues to fail or image quality remains uncertain, escalation is the correct and safest action. Clear CCR documentation helps protect patients, staff, and the service history of the scanner.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide