What This Guide Helps With

Troubleshooting MAGNETOM Skyra patient table movement, docking, positioning, or drive-related faults before escalation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

If a patient is on the table and movement is abnormal, delayed, noisy, or unavailable, stop table operation immediately.

Communicate with MRI staff and assist with safe patient removal using site-approved MRI safety procedures.

Expected outcome: The patient is not moved using a table system that may be unreliable.

Confirm the Reported Table Fault

Ask staff what occurred:

Expected outcome: The issue is clearly identified before troubleshooting begins.

Verify MRI Room Safety Conditions

Confirm that the scan room is safe to enter and that all tools, test items, and personal items are MRI-safe or kept outside Zone IV as required.

Expected outcome: Troubleshooting does not introduce projectile or safety risk.

Check for Patient, Accessory, or Obstruction Interference

Inspect the table path, dock area, side rails, coil cables, pads, straps, sheets, and patient accessories for anything blocking movement.

Expected outcome: The table path and docking area are clear.

If an obstruction is found and removed, retest table movement. If normal operation returns, stop troubleshooting.

Check Emergency Stop and Table Lock Conditions

Verify that any emergency stop, table stop, or movement inhibit condition has been cleared according to site procedure.

Confirm with MRI staff whether an emergency stop was activated during the event.

Expected outcome: Table movement is not being blocked by a safety stop condition.

Verify Table Docking Position

Inspect whether the patient table is fully seated, aligned, and properly docked at the magnet interface.

Look for partial docking, uneven table position, or physical misalignment.

Expected outcome: The table is mechanically aligned before drive movement is tested.

Check Table Communication and Indicator Behavior

Observe table control indicators, system messages, and host console alerts during a controlled movement attempt.

Note whether the system recognizes the table, reports a docking fault, or shows a drive/positioning error.

Expected outcome: The fault is narrowed to movement, docking, recognition, or position feedback.

Test Movement Without a Patient

With MRI staff approval, test table movement in a controlled condition with no patient present.

Use normal table controls and observe for:

Expected outcome: The table moves smoothly through the expected range without abnormal sound or error.

If movement is restored and no errors return, document the finding and return the unit only after operational checks meet site policy.

Check External Cables and Connectors

Inspect accessible table-related cables, docking contacts, and external connection points for looseness, damage, debris, fluid contamination, or strain.

Do not force connectors or disassemble internal table components.

Expected outcome: External communication and power paths appear secure and undamaged.

Inspect for Physical Damage or Fluid Contamination

Check the table surface, base, dock area, drive track area, and accessible exterior panels for signs of impact, cracks, bent components, fluid intrusion, or unusual residue.

Expected outcome: No visible damage or contamination is present that could affect movement.

Check for Unusual Heat, Smell, or Noise

During limited testing, stop immediately if there is burning smell, unusual heat, repeated clicking, grinding, or motor strain.

Expected outcome: The table does not show signs of drive motor, brake, power, or mechanical failure during external troubleshooting.

Power Cycle Only if Appropriate

If MRI operations allow, have authorized MRI staff perform a controlled system/table reset or restart according to site procedure.

Do not interrupt scanning operations or restart the system without MRI staff coordination.

Expected outcome: Temporary software or communication faults clear after restart.

If the table fault clears and controlled movement checks pass, document the event and final status.

Verify Final Operational Status

Before returning the system to use, confirm:

Expected outcome: The MAGNETOM Skyra table is safe and reliable for clinical use.

If the Problem Persists

If table movement, docking, position feedback, or drive faults continue after external checks, the likely cause may involve internal drive components, sensors, brakes, table electronics, docking hardware, or system communication.

The MRI system or affected patient table should be removed from service, labeled Out of Service, and sent for Siemens-qualified service, repair, or bench evaluation.

Knowing when to stop is proper troubleshooting. Do not continue repeated table movement attempts if there is abnormal resistance, noise, heat, smell, or unreliable movement.

Clinical Use Tip

Do not troubleshoot table drive or docking faults with an active patient on the table unless the immediate goal is safe patient removal. Move the patient to another safe option first, then troubleshoot only when MRI staff confirm it is safe.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"MRI staff reported the Siemens MAGNETOM Skyra patient table would not dock fully and displayed a table movement fault at the console."

Cause

What was observed during troubleshooting.

Example:
"Found the docking area obstructed by a coil cable routed across the table interface, preventing full table seating."

Resolution

What action was taken.

Example:
"Removed the obstruction, verified table docking and in/out movement without a patient, confirmed no further table movement faults, and returned the system to service."

Helpful Details to Include (If Known)

Final Thought

Patient table faults on an MRI system should be handled carefully because they affect patient access, scanner availability, and safety. Start with simple external causes such as obstructions, docking alignment, emergency stop conditions, and visible cable issues. If the fault remains or movement is unreliable, stop and escalate for qualified repair. Clear CCR documentation helps show what was checked and why the device was returned to service or removed.

That is successful troubleshooting.