Siemens Healthineers MAGNETOM Sola

Numaris X / Host / Reconstruction / Startup Fault

On this page

Asset Type

MRI System

Manufacturer

Siemens Healthineers

Model

MAGNETOM Sola

What This Guide Helps With

Troubleshooting Numaris X startup, host, reconstruction, or workstation faults before escalating MRI system service.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm no patient is inside the bore and no exam is actively running before troubleshooting the MAGNETOM Sola.

If the system fault occurs during patient care, stop the exam only when clinically safe, remove the patient from the scanner room, and have staff use another available MRI system or approved downtime process.

Expected outcome: The patient is safe, the scan room is controlled, and troubleshooting is not performed during active scanning.

Confirm the Exact Fault and System State

At the Numaris X console, document the exact message, error code, screen behavior, and when the fault occurs.

Check whether the issue appears during:

Expected outcome: The issue is clearly identified as a startup, host, reconstruction, or workflow fault.

Why it matters: Siemens Numaris X systems use system messages and message history to summarize software, scanner, and MR View&GO-related issues, so exact message capture is important before escalation.

Check the Console and Workstation Power Condition

Verify the operator console, monitors, keyboard, mouse, and any connected workstation components are powered normally.

Look for:

Expected outcome: The issue is not caused by a simple console display, input, or local power problem.

If reseating an external power or display connection restores normal operation, confirm successful startup and stop.

Check for Facility Power or UPS Issues

Confirm whether the MRI console room, equipment room, and associated workstation power sources are stable.

Ask whether there was a recent power outage, emergency power transfer, electrical work, UPS alarm, or breaker event.

Expected outcome: External power instability is either identified or ruled out.

If power instability is suspected, notify Facilities and do not continue repeated restart attempts until the power source is verified.

Review System Messages and Message History

Open the system message area or message history if the system allows access.

Document any messages related to:

Expected outcome: The fault category is better defined for service documentation.

Siemens identifies Numaris X system messages as being summarized in message history, with different message areas depending on single- or dual-monitor configuration.

Ask Clinical Staff What Changed Before the Fault

Check whether the issue followed:

Expected outcome: A likely trigger is identified or ruled out.

If the issue began after a known software update or vendor service action, document that clearly before escalation.

Check Local Storage and Reconstruction Behavior

If the system starts but reconstruction fails, verify whether the issue affects one exam, one protocol, or all studies.

Ask the technologist whether:

Expected outcome: The issue is separated into a single-study reconstruction problem or a broader host/reconstruction system fault.

If only one exam or sequence is affected, preserve the study data and escalate to Siemens support with the exam details.

Check Network-Dependent Functions

Determine whether the fault is truly host/reconstruction related or actually a network/archive issue.

Verify whether the MAGNETOM Sola can:

Expected outcome: PACS, worklist, and network transfer issues are not mistaken for a host or reconstruction failure.

If other imaging systems are also having PACS or worklist problems, escalate through imaging IT or network support rather than treating this as an isolated MRI scanner failure.

Check External Network Connections in the Console Area

Visually inspect accessible network connections at the console, workstation area, and approved external patch points.

Do not enter restricted MRI equipment cabinets or disturb vendor-controlled internal connections.

Expected outcome: No obvious disconnected external network cable or patch issue is found.

If a loose external network connection is corrected and communication returns, verify normal workflow with clinical staff and stop.

Perform an Orderly Restart Only if Safe and Appropriate

If no patient is active and the system is in a safe state, perform only the approved shutdown or restart process available to the site.

Do not force power-cycle MRI cabinets, reconstruction hardware, or equipment room components unless directed by Siemens service or site policy.

Expected outcome: The system restarts cleanly and startup errors clear.

Siemens Healthineers has Numaris X shutdown/restart guidance for MAGNETOM systems, including Sola, and notes that the host computer and MaRS are included in the computer system shutdown process.

If the restart clears the issue, confirm login, patient browser access, test workflow readiness, and image transfer status before returning the system to clinical use.

Verify the System Reaches a Clinically Ready State

After startup, confirm with MRI staff that the system reaches the normal home screen and is ready for standard workflow.

Check for:

Expected outcome: The system is not returned to service based only on a cleared screen; basic operational readiness is confirmed.

Do Not Clear, Delete, or Modify Study Data Without Approval

If storage, reconstruction queue, or patient browser problems are suspected, do not delete exams, raw data, or reconstruction files unless directed by Siemens service and approved by site policy.

Expected outcome: Patient data is protected and vendor support can review the fault condition.

Escalate Repeated or Persistent Startup/Reconstruction Faults

If the MAGNETOM Sola repeatedly fails startup, freezes, cannot access host functions, cannot reconstruct images, or shows recurring MaRS/reconstruction faults, remove it from clinical use.

Expected outcome: The system is treated as a service-level MRI failure, not a routine user workflow issue.

If the Problem Persists

If power, monitor/input devices, external network conditions, message review, workflow checks, and an orderly restart do not resolve the issue, the likely cause is internal to the host, reconstruction system, software environment, storage/database, or scanner communication path.

The MAGNETOM Sola should be:

Knowing when to stop is proper troubleshooting. MRI host and reconstruction faults can affect exam completion, image availability, and patient data integrity.

Clinical Use Tip

Do not troubleshoot Numaris X startup or reconstruction faults while a patient is on the table or inside the bore. Get the patient safely off the system first, then troubleshoot only when the scan room and console area are controlled.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"MRI staff reported the Siemens MAGNETOM Sola would not complete Numaris X startup and displayed a host/reconstruction fault at the operator console."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering verified console power, monitor/input operation, external network connections, and system message history; fault returned after an orderly restart, indicating a likely internal host or reconstruction system issue."

Resolution

What action was taken.

Example:
"Removed the MRI system from service, labeled it Out of Service, notified MRI leadership, and escalated to Siemens Healthineers service with documented fault messages and startup behavior."

Helpful Details to Include (If Known)

Final Thought

MAGNETOM Sola startup and reconstruction faults should be approached carefully because they may affect image creation, workflow continuity, and patient record availability. Start with safe, external, verifiable checks, document the exact system behavior, and escalate when the fault points beyond Clinical Engineering troubleshooting.

That is successful troubleshooting.