Skytron 6500 Series

Hand Control or Pendant Failure

On this page

Asset Type

Surgical Table

Manufacturer

Skytron

Model

6500 Series

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting a Skytron 6500 Series surgical table when the hand control or pendant does not operate the table. Common symptoms include:

Focus is on external, verifiable checks before considering internal repair.

Step-by-Step Troubleshooting

Visual Inspection

Check Power and Table Status

Cable Continuity Test

Battery Check (if applicable)

Test Table Response

If the Problem Persists

Common external causes have been ruled out. The issue is likely internal to the table’s control electronics or the pendant itself.

Knowing when to stop avoids unnecessary risk or further damage.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“OR staff report that pressing buttons on the hand control or pendant does not move the table.”

Cause

What was observed during troubleshooting.

Example:
“Visual inspection showed intact cabling; pendant does not respond; table powers on normally.”

Resolution

What action was taken.

Example:
“Hand control/panel swapped with known working unit for verification. Table removed from service and sent for bench evaluation due to likely internal electronics failure.”

Helpful Details to Include

Final Thought

Patient safety and methodical troubleshooting are paramount. Start with simple checks like connections and batteries, confirm table status, and escalate appropriately when internal failures are suspected. Document all steps clearly for future reference.

That is successful troubleshooting.