Skytron 6500 Series

Hydraulic Leaks or Slow Positioning

On this page

Asset Type

Surgical Table

Manufacturer

Skytron

Model

6500 Series

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting a Skytron 6500 Series surgical table that exhibits hydraulic issues, such as slow positioning, uneven movements, or visible fluid leaks. These issues may impact table safety and surgical workflow. Focus is on external and easily verifiable checks before assuming internal hydraulic failure.

Common symptoms include:

Step-by-Step Troubleshooting

Ensure Patient Safety First

Check Power and Setup

Inspect Hand Control and Pendant

Check Hydraulic Fluid Level (External Reservoir if Accessible)

Visual Inspection of Hydraulic Lines and Connections

Observe Table Movement Under No Load

Check for Obstructions

If the Problem Persists

If table continues to move slowly, drift, or shows persistent hydraulic leaks after external checks:

Clinical Engineering should avoid internal cylinder or pump disassembly; this requires manufacturer service.

Knowing when to escalate prevents unsafe operation and further damage.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
β€œTable moves slowly during height and tilt adjustments; visible hydraulic fluid under base.”

Cause

What was observed during troubleshooting.

Example:
External inspection revealed minor leaks at a cylinder fitting and low fluid in reservoir.

Resolution

What action was taken.

Example:
Table removed from service, labeled Out of Service, and sent for authorized repair.

Helpful Details to Include

Final Thought

Hydraulic performance is critical for safe surgical table operation. External checks allow Clinical Engineering to rule out simple issues, but persistent slowness or leaks always warrant professional repair. Proper documentation ensures clear communication and maintains patient safety.

That is successful troubleshooting.