SMITHS MEDICAL CADD LEGACY

BATTERY WON’T HOLD CHARGE / WON’T POWER ON

On this page

Asset Type

Infusion Pump

Manufacturer

Smiths Medical

Model

CADD Legacy

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting a Smiths Medical CADD Legacy infusion pump that runs on two AA batteries and either won’t power on or fails to operate correctly. Common symptoms include:

The guide emphasizes external checks first, avoiding internal electronics work.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Why: Troubleshooting should never interrupt active patient therapy.

Verify Battery Type and Condition

Why: Weak or incorrect batteries are the most common cause of power failure.

Check Battery Orientation and Contacts

Why: Poor contact prevents the pump from powering on even with good batteries.

Attempt Power-On With Fresh Batteries

Why: Confirms whether the issue is battery-related or internal.

Observe LED Indicators or Error Messages

Why: Indicates if pump electronics recognize the batteries.

Reset Pump Electronics

Why: A simple power-cycle can clear minor electronic faults.

If the Problem Persists

Fresh batteries inserted correctly, contacts cleaned, and correct polarity verified.

Device still will not power on.

Stopping at this point is correct troubleshooting once external causes have been ruled out.

Clinical Use Tip

Never troubleshoot on an actively infusing patient.

Move the patient to a backup pump first.

Only handle battery issues when the device is safely disconnected from patient therapy.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump will not power on; batteries appear dead.”

Cause

What was observed during troubleshooting.

Example:
“Two AA batteries replaced with fresh alkalines; correct polarity and contacts verified; pump still fails to power on.”

Resolution

What action was taken.

Example:
“Device removed from service and sent to Biomedical Engineering for bench evaluation.”

Helpful Details to Include (If Known)

Final Thought

Always start with simple external checks—fresh batteries and clean contacts—before assuming internal failure. Proper documentation and knowing when to escalate ensures patient safety and preserves pump functionality.

That is successful troubleshooting.