SMITHS MEDICAL CADD LEGACY

CASSETTE NOT SEATED / “CHECK CASSETTE” ERROR

On this page

Asset Type

Infusion Pump

Manufacturer

Smiths Medical

Model

CADD Legacy

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting a Smiths Medical CADD Legacy infusion pump that displays a “Check Cassette” error or refuses to start infusion. Symptoms may include:

Focus is on external checks for cassette seating and latch engagement before considering internal repair.

Step-by-Step Troubleshooting

Power Cycle the Pump

Why: Clears temporary sensor misreads and ensures error isn’t a transient fault.

Inspect the Cassette and Slot

Why: Even minor blockage can prevent proper seating.

Reinsert the Cassette Properly

Why: Proper engagement is required for the sensor to detect the cassette.

Check the Latch Mechanism

Why: Latch interference can prevent sensor detection.

Verify Sensor Engagement

Attempt Infusion Only After Error Clears

If the Problem Persists

External causes (cassette seating, debris, latch issues) have been ruled out. Internal latch or sensor failure is likely.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Pump displays “Check Cassette” error and will not start infusion.”

Cause

What was observed during troubleshooting.

Example:
“Cassette partially seated or latch obstructed by debris.”

Resolution

What action was taken.

Example:
“Cleared debris, reinserted cassette fully, verified latch and sensor engagement, pump operational.”

Helpful Details to Include

Final Thought

Troubleshooting starts with simple, external verification of cassette seating and latch engagement. Escalate only when internal sensor or latch mechanisms fail, always keeping patient safety as the top priority. Clear CCR documentation preserves accountability and supports maintenance decisions.

That is successful troubleshooting.

Related Guides

Was this guide helpful?