What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting a Smiths Medical CADD Legacy infusion pump when reported infusion rates appear incorrect or when staff report programming errors. Symptoms may include unexpected infusion behavior, incorrect rate entry, infusion stopping prematurely, or the pump refusing programmed settings.

The goal is to verify user setup, cassette configuration, and programming parameters before assuming a device malfunction.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the reported issue with the clinical user

Check the infusion mode selected on the pump

Verify programmed parameters

Confirm the correct cassette or medication reservoir is installed

Power cycle the pump

Test programming with a known standard configuration

Inspect the keypad for proper operation

Check for error messages or alerts

Verify battery condition

If the Problem Persists

If the pump continues to accept incorrect programming, fails to operate at the programmed rate, or behaves inconsistently after these checks, common external causes have been ruled out.

The issue may involve internal components such as keypad membrane failure, internal control board issues, software malfunction, or pump drive mechanism problems.

At this point the device should be removed from service, labeled Out of Service, and sent for repair or further bench evaluation.

Knowing when to escalate prevents unnecessary risk and is an important part of proper troubleshooting.

Clinical Use Tip

Never troubleshoot infusion pump programming while the pump is connected to a patient.

If infusion accuracy is in question, immediately transition the patient to another verified pump before beginning any troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Nursing reported that the CADD Legacy infusion pump appeared to deliver medication faster than the programmed rate and programming values did not match expected infusion behavior.”

Cause

What was observed during troubleshooting.

Example:
“During inspection, pump programming was verified and keypad testing revealed intermittent input errors when entering numerical values.”

Resolution

What action was taken.

Example:
“Pump was removed from service, labeled Out of Service, and sent for repair due to suspected keypad malfunction.”

Helpful Details to Include (If Known)

Final Thought

When infusion behavior appears incorrect, many issues originate from programming configuration, cassette setup, or keypad input errors rather than internal pump failure. Systematically verifying external factors first ensures safe operation, prevents unnecessary repairs, and supports accurate documentation of the issue.

That is successful troubleshooting.