Smiths Medical CADD Legacy

Remote Dose Cord or Patient Bolus Button Does Not Respond

On this page

Asset Type

Infusion Pump

Manufacturer

Smiths Medical

Model

CADD Legacy

What This Guide Helps With

Troubleshoots an unresponsive remote dose control caused by connection, accessory damage, configuration, programming state, controls, or a pump interface problem.

Step-by-Step Troubleshooting

1. Protect the Patient and Verify Therapy Continuity

If patient-controlled dosing is clinically required and the remote dose control is unreliable, provide an alternate verified pump or approved method of administering ordered therapy.

Do not leave a patient dependent on a nonfunctional dose control.

Expected outcome: Required therapy remains safely available while the accessory is evaluated.

2. Confirm the Exact Complaint

Determine whether the remote button:

Record any display or alarm response when the button is pressed.

Expected outcome: The failure is clearly distinguished between accessory input, pump response, and programmed-dose behavior.

3. Inspect the Remote Dose Accessory

Remove the device from patient use as appropriate and visually inspect the remote dose cord and button.

Look for:

Do not repair a damaged patient-accessible cord with tape or other temporary methods.

Expected outcome: The accessory is visibly intact or is identified as damaged and removed from use.

4. Inspect and Reseat the Connection

Inspect the pump's accessible remote-control connection for contamination, damage, or a loose fit.

Reconnect the accessory securely without forcing the connector.

Expected outcome: The connection is fully seated and mechanically stable.

5. Verify the Pump Is Configured to Accept the Requested Dose

Have qualified clinical personnel confirm that the programmed therapy permits the expected patient dose and that the pump is in an operating state where the control should be active.

Do not alter clinical programming simply to make the accessory appear functional.

Expected outcome: The pump is confirmed to be in a valid state for dose-control testing.

6. Substitute a Known-Good Remote Dose Accessory

Connect a compatible known-good remote dose cord or patient control intended for the device.

Test using an approved nonpatient setup.

Expected outcome: If the known-good accessory responds consistently, the original accessory is defective and should be replaced. Troubleshooting can stop after final pump verification.

7. Check for Cable-Position Intermittency

During bench testing only, gently move the accessory cable and connector through normal handling positions.

Do not sharply bend, pull, or stress the connection.

Expected outcome: Normal movement does not interrupt recognition or response.

8. Verify Pump Response and Delivery Function

Using approved test procedures and suitable test equipment, confirm that a valid dose request is recognized appropriately and that the pump's response is consistent with its authorized programming.

Do not perform patient-dose testing using an actual patient connection.

Expected outcome: The pump recognizes the control consistently and functions normally in the approved test setup.

9. Determine Whether the Pump Can Return to Service

If the problem was isolated to the removable accessory and the pump passes functional testing with a known-good accessory, replace the defective accessory and document the result.

If multiple known-good accessories fail or the pump connection is loose or damaged, remove the pump from service.

Expected outcome: Only a verified pump-accessory combination returns to clinical use.

If the Problem Persists

After connection, accessory condition, programming state, and known-good accessory substitution have been ruled out, the problem may involve the pump's accessory interface, control-input circuitry, configuration, connector assembly, or another service-level condition.

The device should be:

After repair, verify the remote-control input, normal pump operation, alarm functions, and required delivery performance before return to service.

Recognizing that repeated failure with known-good accessories requires escalation is proper troubleshooting.

Clinical Use Tip

If patient-controlled dosing is part of the prescribed therapy, verify the complete pump-and-button pathway before returning the system to the patient.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Nursing reported that the patient bolus button intermittently failed to respond when pressed."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found an intermittent remote dose cord that failed when the cable was moved near its strain relief."

Resolution

What action was taken.

Example:
"The defective remote dose cord was replaced, and the pump was tested with a known-good control to verify consistent dose-request recognition and normal operation."

Helpful Details to Include (If Known)

Final Thought

Begin with the removable remote-dose accessory, connection, and programmed operating state before suspecting the pump itself. Confirm the entire input pathway under controlled conditions and escalate any pump-side intermittency.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide