Smiths Medical Medfusion 3500

Battery Charging Failure / Won’t Power On

On this page

Asset Type

Infusion Pump

Manufacturer

Smiths Medical

Model

Medfusion 3500

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting a Medfusion 3500 syringe infusion pump that will not power on or does not appear to charge. Symptoms may include a blank display, the pump shutting off when unplugged, or the battery not charging while connected to AC power. The guide focuses on verifying external power sources and easily correctable conditions before assuming internal battery or power supply failure.

Step-by-Step Troubleshooting

Verify the AC power source

Inspect the power cord and adapter

Check the power indicator lights

Attempt a power cycle

Inspect the charging connector

Allow the pump to charge

Check for battery-related symptoms

If the Problem Persists

If external power sources, cables, and connectors have been verified and the pump still will not power on or charge, the issue is likely related to an internal battery failure, charging circuit fault, or internal power supply issue.

At this point the device should be:

Knowing when external causes have been ruled out and escalation is appropriate is an important part of effective Clinical Engineering troubleshooting.

Clinical Use Tip

Never troubleshoot an infusion pump that is actively connected to a patient. Ensure the patient has been safely transferred to another functioning pump before beginning any troubleshooting steps.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Medfusion 3500 infusion pump would not power on and appeared completely dead.”

Cause

What was observed during troubleshooting.

Example:
“Pump did not power on during testing. Verified outlet was functional and power supply was connected. Pump only briefly illuminated when plugged in and shut off immediately when unplugged, indicating battery failure.”

Resolution

What action was taken.

Example:
“Pump removed from service, labeled Out of Service, and sent for battery replacement and internal power circuit evaluation.”

Helpful Details to Include (If Known)

Final Thought

Many infusion pump power issues are caused by simple external factors such as faulty outlets or damaged power cords. Systematically verifying these first prevents unnecessary repair actions. When those checks are complete and the pump still fails to charge or power on, removing the device from service and documenting the findings ensures both patient safety and efficient repair follow-up.

That is successful troubleshooting.