On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshoots abnormal drive movement caused by syringe loading, obstruction, excessive resistance, mechanical interference, or service-level drive problems.
Step-by-Step Troubleshooting
1. Protect the Patient and Maintain Therapy
If the drive is not advancing reliably, stop using the affected pump for patient therapy and transfer the infusion to another verified device as clinically appropriate.
Expected outcome: Patient therapy is maintained without dependence on an unreliable drive mechanism.
2. Confirm the Drive Symptom
Determine whether the drive head does not move, stops during infusion, slips against the syringe plunger, moves intermittently, or makes abnormal mechanical sounds. Record when the failure occurs.
Expected outcome: The reported drive behavior is clearly defined.
3. Inspect Syringe Loading
Remove and reinstall the syringe, checking that the barrel, flange, and plunger are correctly seated and aligned. Verify that no tubing, labels, or external objects interfere with drive travel.
Expected outcome: The syringe is properly aligned with an unobstructed drive path.
4. Inspect the Syringe
Check for a binding or damaged plunger, deformed barrel, contamination, or unusual resistance. Replace the test syringe with an approved known-good syringe if anything appears abnormal.
Expected outcome: A mechanically normal syringe is available for testing.
5. Inspect the External Drive Path
With the pump removed from clinical service, inspect accessible areas along the drive path for debris, dried fluid, foreign material, visible damage, or obstruction. Do not open the drive housing.
Expected outcome: No external obstruction is present.
6. Observe External Drive Movement
Operate the drive only as permitted during normal loading or approved bench testing. Watch for hesitation, uneven motion, binding, or inability to maintain contact with the syringe plunger.
Expected outcome: Drive movement is smooth and consistent without slipping or stalling.
7. Test With a Known-Good Syringe and Setup
Install a known-good compatible syringe with a simple bench setup and repeat operation. Avoid using a restrictive clinical line or accessory that could introduce excessive downstream resistance during this comparison.
Expected outcome: The drive advances normally with the known-good setup. If so, the original syringe or external fluid-path resistance was the likely cause.
8. Evaluate External Flow Resistance
Inspect connected tubing, clamps, valves, stopcocks, filters, and other accessories for restrictions that may create excessive resistance. Do not assume a drive defect solely because advancement stopped under a restrictive setup.
Expected outcome: The external fluid path is patent and properly configured.
9. Perform Functional Verification
Using appropriate infusion-pump test equipment, verify that the pump drives the syringe consistently through an appropriate functional test without stalls, slipping, abnormal noises, or interruption.
Expected outcome: Drive operation remains stable throughout testing. If successful, complete applicable return-to-service verification.
10. Escalate Persistent Drive Problems
If the drive stalls, slips, binds, fails to advance, or produces abnormal mechanical behavior with a known-good syringe and unrestricted bench setup, remove the pump from service.
Expected outcome: A pump with a potentially unreliable drive system is sent for qualified service evaluation.
If the Problem Persists
Common external causes including syringe loading, plunger resistance, tubing restriction, and drive-path obstruction have been ruled out. Remaining possibilities include internal mechanical drive wear, alignment problems, motor or feedback-related conditions, or other service-level faults.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate manufacturer documentation and approved test equipment
- Repaired or configured only by qualified personnel
Complete appropriate performance testing following repair before return to clinical use. Recognizing that persistent drive instability exceeds safe external troubleshooting is proper troubleshooting.
Clinical Use Tip
Do not leave a patient dependent on a syringe pump whose drive motion is intermittent, slipping, or stalled, even if restarting temporarily restores movement.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Nursing reported that the Medfusion 3500 stopped advancing the syringe during an infusion."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found a partially closed clamp in the test setup creating excessive downstream resistance; drive operation was normal with an unrestricted line."
Resolution
What action was taken.
Example:
"The restricted setup was corrected, the pump completed functional testing normally, and the unit was returned to service."
Helpful Details to Include (If Known)
- Syringe type and size
- Point during infusion when drive stopped
- Any abnormal noise
- Plunger condition
- External tubing and clamp condition
- Known-good syringe result
- Whether drive movement was intermittent
- Infusion analyzer results
- Condition before and after correction
- Final device status
- Syringe type and size
- Point during infusion when drive stopped
- Any abnormal noise
- Plunger condition
- External tubing and clamp condition
- Known-good syringe result
- Whether drive movement was intermittent
- Infusion analyzer results
- Condition before and after correction
- Final device status
Final Thought
Evaluate syringe loading and external resistance before condemning the drive system. Persistent abnormal drive movement with a known-good setup requires removal from service, controlled testing, and clear documentation.
That is successful troubleshooting.