Smiths Medical Medfusion 3500

Occlusion or Line Blockage Alarms

On this page

Asset Type

Infusion Pump

Manufacturer

Smiths Medical

Model

Medfusion 3500

What This Guide Helps With

This guide assists Clinical Engineering in troubleshooting occlusion or line blockage alarms on a Medfusion 3500 syringe infusion pump. These alarms occur when the pump detects excessive pressure in the infusion line, indicating that fluid flow may be restricted.

Most occlusion alarms are caused by external issues such as tubing problems, closed clamps, or downstream flow restrictions rather than internal pump failure. This guide focuses on verifying these common causes before assuming a pump malfunction.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Confirm the Alarm Message

Check for Closed Clamps in the Infusion Line

Inspect the Tubing Path for Kinks or Compression

Check the Syringe Installation

Verify Correct Syringe Type and Size

Inspect the Pressure Sensor Area

Replace the Tubing Set or Syringe

Restart the Pump and Observe Operation

If the Problem Persists

If occlusion alarms continue after verifying clamps, tubing, syringe installation, and replacing disposable components, the issue may involve:

At this point the device should be:

Recognizing when external causes have been ruled out is an important part of proper troubleshooting.

Clinical Use Tip

Never troubleshoot infusion pumps while they are actively delivering medication to a patient.

Always ensure the infusion is safely transferred to another pump before removing the device from service.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
“Nursing reported Medfusion 3500 pump repeatedly alarming for ‘Occlusion’ during infusion.”

Cause

What was observed during troubleshooting.

Example:
“Inspection revealed IV tubing kinked under patient bed rail causing downstream flow restriction.”

Resolution

What action was taken.

Example:
“Tubing repositioned and infusion restarted. Pump tested and operated normally without further occlusion alarms.”

Helpful Details to Include (If Known)

Final Thought

Occlusion alarms are designed to protect patients by detecting restricted medication flow. Most alarms are caused by simple setup or tubing issues rather than pump failure. A logical approach—starting with clamps, tubing, and syringe setup—helps resolve many issues quickly while ensuring patient safety. When external causes have been ruled out, removing the device from service for repair is the appropriate action.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide