Smiths Medical Medfusion 4000

Alarm: Calibration Required

On this page

Asset Type

Infusion Pump

Manufacturer

Smiths Medical

Model

Medfusion 4000

What This Guide Helps With

This guide helps troubleshoot Calibration Required alarms on the Smiths Medical Medfusion 4000 infusion pump. This alarm indicates the pump may not be safe for normal infusion use until calibration, maintenance, or verification checks are completed. Calibration-related conditions may affect syringe recognition, plunger position, force sensing, or delivery accuracy.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error/Alarm and Behavior

Remove the Pump From Clinical Use

Check for Recent Damage or Drop History

Verify Power and Startup Behavior

Check External Condition Before Calibration

Confirm Whether Annual Maintenance Is Due

Perform Required Calibration or Calibration Verification

Perform Functional Checks After Calibration

Return the Pump to Service Only After Passing Checks

If the Problem Persists

If the Calibration Required alarm remains after required calibration or calibration verification, external causes and routine maintenance causes have been ruled out.

A persistent calibration-related alarm may indicate a failed or drifting sensor, syringe size detection issue, plunger position issue, force sensor issue, mechanical alignment problem, or internal device fault.

Remove the device from service, label it Out of Service, and send it for repair or further bench evaluation.

Do not repeatedly clear the alarm and return the pump to clinical use.

Clinical Use Tip

Never troubleshoot or calibrate a syringe pump while it is connected to an active patient. Move the patient to a backup device first, then begin calibration-related troubleshooting or bench evaluation.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Medfusion 4000 reported Calibration Required alarm at startup. Pump was removed from clinical use because calibration status could affect normal infusion reliability."

Cause

What was observed during troubleshooting.

Example:
"Annual calibration verification was due. No external damage or syringe loading obstruction found during inspection."

Resolution

What action was taken.

Example:
"Removed pump from service, performed required annual maintenance and calibration verification through Biomed Calibration functions. Pump passed post-calibration functional checks and was returned to service."

Helpful Details to Include (If Known)

Final Thought

A Calibration Required alarm should be treated as a safety stop, not a routine user alarm. Clinical Engineering should verify the alarm, remove the pump from service, check for damage or overdue maintenance, perform the required calibration or calibration verification, and only return the pump after it passes required checks. Clear documentation protects the patient, the department, and the repair history.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide