Smiths Medical Medfusion 4000

Alarm: Occlusion - Check Infusion Line

On this page

Asset Type

Infusion Pump

Manufacturer

Smiths Medical

Model

Medfusion 4000

What This Guide Helps With

This guide helps troubleshoot the Occlusion - Check Infusion Line alarm on a Smiths Medical Medfusion 4000 infusion pump. This alarm indicates that the force sensor in the plunger driver sensed an occlusion because the force value exceeded the configured limit.

The condition may be caused by a closed clamp, kinked tubing, blocked extension set, catheter resistance, high pressure setup, syringe loading issue, inappropriate occlusion configuration, or force sensing problem.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check the Infusion Line for Simple External Causes

Check the Syringe and Pump Loading

Check for Catheter or Patient-Side Resistance

Review the Infusion Setup, Fluid, and Rate

Check Occlusion Alarm Configuration

Consider FlowSentry Pressure Monitoring if Used

Check for Mechanical Binding or Physical Damage

Check Force Sensor Calibration if the Alarm Recurs

Stop When Resolved

If the Problem Persists

If the Occlusion - Check Infusion Line alarm continues after external causes, line setup, catheter resistance, syringe loading, configuration, and force sensor calibration have been ruled out, the issue may involve the force sensing system or plunger drive mechanism.

Remove the pump from service.

Label it Out of Service.

Send it for repair or bench evaluation.

Persistent occlusion alarms may indicate force sensor failure, plunger driver mechanical binding, plunger mechanism damage, internal contamination, calibration failure, plunger PCB issue, or Main PCB signal processing issue.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device first if infusion delivery is interrupted or the occlusion alarm cannot be safely resolved at the bedside.

Do not repeatedly restart an infusion after an active occlusion alarm without confirming line patency, catheter condition, and medication delivery safety.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Medfusion 4000 infusion pump reported Occlusion - Check Infusion Line during infusion. User stated pump stopped delivery and alarm returned after restart."

Cause

What was observed during troubleshooting.

Example:
"Force sensor detected pressure above configured occlusion limit. Infusion line was found kinked near the patient connection, causing downstream restriction."

Resolution

What action was taken.

Example:
"Removed kink from infusion line, verified syringe was loaded correctly, confirmed alarm cleared, and ran pump without repeated occlusion alarm. Device returned to service."

Helpful Details to Include (If Known)

Final Thought

Occlusion alarms should be treated as possible therapy interruptions until proven otherwise. Start with the infusion path, syringe loading, catheter resistance, clinical setup, and configuration before suspecting device failure. If the alarm continues after common causes are ruled out, approved force sensor calibration or repair escalation may be needed. Clear CCR documentation helps show whether the issue was clinical setup, configuration, calibration, or a true device fault.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide