What This Guide Helps With
Troubleshooting Xprezzon monitors not communicating with the central station, network, remote viewing, or alarm display systems.
Step-by-Step Troubleshooting
Ensure Patient Safety First
If the Xprezzon is actively monitoring a patient and communication to the central station is lost, notify clinical staff immediately.
Do not assume central station alarm visibility is active until confirmed.
Move the patient to another monitored bed, use a known-good monitor, or provide direct bedside observation according to clinical policy.
Expected outcome: Patient monitoring and alarm awareness are maintained without relying on failed network communication.
Why it matters: Loss of network connectivity may prevent alarms and patient data from appearing at central stations or other networked displays. Spacelabs documentation identifies Xprezzon as a network-capable bedside monitor, and FDA recall documentation has described loss of network connectivity as affecting central station alarm and data display.
Confirm the Reported Communication Failure
Check whether the issue is:
- No patient data at the central station
- Bedside monitor missing from the central station
- Delayed or frozen waveform display
- Alarm not appearing centrally
- Remote view unavailable
- Monitor visible locally but not over the network
Confirm whether bedside monitoring functions still display correctly on the Xprezzon.
Expected outcome: The problem is identified as a network or central station communication issue, not a general monitor failure.
Check Whether the Issue Is Isolated or Widespread
Ask staff whether other Spacelabs monitors on the same unit are communicating normally.
Check nearby monitors connected to the same network area or central station.
Expected outcome: One failed monitor suggests a local cable, port, configuration, or monitor issue. Multiple failed monitors suggest a network switch, central station, server, or IT infrastructure issue.
If multiple monitors are affected, escalate to Clinical Engineering leadership and IT/network support before replacing parts.
Verify Bedside Monitor Power and Normal Operation
Confirm the Xprezzon is powered on, fully booted, and not frozen.
Check for normal waveform and numeric display at the bedside.
Confirm there are no obvious system errors, lockups, repeated restarts, or abnormal boot behavior.
Expected outcome: The monitor is operating locally before network communication is investigated.
If the monitor is frozen or repeatedly rebooting, remove it from service and send for bench evaluation.
Check Network Cable Connection
Inspect the network cable at the monitor and wall jack.
Confirm the cable is fully seated and not loose, bent, crushed, pulled tight, or routed where bed movement may strain it.
Reseat both ends of the cable if accessible and safe.
Expected outcome: A loose or partially disconnected network cable is corrected.
If communication restores, verify central station visibility and stop.
Inspect Network Link Indicators
Check the network port on the monitor, dock, wall jack, or associated network adapter if visible.
Look for link/activity lights.
Expected outcome: Link lights should indicate an active physical network connection.
If no link light is present, continue checking the cable, wall port, and network pathway.
Swap with a Known-Good Network Cable
Replace the current network cable with a known-good hospital-approved cable.
Avoid using damaged, unlabeled, excessively long, or non-approved cables.
Expected outcome: Communication returns if the original cable was defective.
If communication restores, document the failed cable and remove it from use. Stop.
Test the Wall Network Jack
Move the monitor to a known-good network jack assigned for patient monitoring, or connect a known-good monitor to the suspect jack if appropriate.
Do not connect to random wall ports unless the network assignment is verified.
Expected outcome: The issue follows either the wall jack/network drop or the monitor.
If the suspect jack fails with multiple monitors, escalate to IT/network support.
If the Xprezzon fails on a known-good jack, continue troubleshooting the monitor-side issue.
Confirm Correct Network Location and Bed Assignment
Verify the monitor is being used in the correct room, bed, care area, or network zone.
Confirm the central station is looking for the correct bed label or monitor assignment.
Check whether the monitor was recently moved, swapped, replaced, or returned from repair.
Expected outcome: The monitor appears under the correct bed or central station location.
Why it matters: A monitor may be working normally but assigned to a different bed, sector, or central station view.
Check Patient Admit / Discharge / Monitor Association
Confirm whether the patient has been admitted, discharged, transferred, or reassigned at the monitor or central station.
Ask clinical staff whether the issue began after a bed move, patient transfer, or monitor swap.
Expected outcome: Patient and bed association is corrected if the monitor was not properly assigned.
If communication restores after correct assignment, verify waveform and alarm visibility at the central station and stop.
Review Central Station Display and Alarm Behavior
At the central station, confirm whether the monitor is:
- Missing completely
- Present but showing no waveforms
- Showing old or frozen data
- Displaying a communication or network message
- Assigned to the wrong bed
- Silenced, hidden, discharged, or removed from view
Expected outcome: The central station side of the failure is confirmed.
If the issue is only at one central station display, troubleshoot that workstation or escalate appropriately.
Restart Only When Clinically Safe
If the patient is not dependent on the affected monitor, or monitoring has been safely transferred, restart the bedside monitor according to facility procedure.
Do not reboot a monitor that is actively needed for continuous patient monitoring without clinical approval.
Expected outcome: Temporary communication faults may clear after a controlled restart.
If communication restores, confirm stable central station display and alarm visibility before returning to use.
Check for Recent Configuration or Network Changes
Ask whether any of the following occurred recently:
- Network maintenance
- Central station restart
- Monitor replacement
- Room renovation
- Wall jack repair
- IP address, VLAN, or switch changes
- Software update
- Bedside monitor moved from another unit
Expected outcome: Recent changes help identify whether the failure is device-specific or infrastructure-related.
Escalate for Network or System-Level Review
If the cable, wall jack, bed assignment, and local monitor operation appear normal, escalate to the appropriate support group.
This may include Clinical Engineering, IT/networking, Spacelabs system administration, or vendor support depending on site structure.
Provide the affected room, bed, monitor serial number, network port, central station name, time of failure, and whether other beds are affected.
Expected outcome: Network configuration, central station routing, switch port status, or system communication issues are reviewed by the proper support team.
Remove from Service if the Monitor Itself Remains Suspect
If the Xprezzon does not communicate on a known-good cable, known-good network jack, and known-good central station assignment, remove it from clinical use.
Label it Out of Service and send it for bench evaluation.
Expected outcome: A potentially faulty monitor network interface or internal communication problem is removed from patient care.
If the Problem Persists
If the Xprezzon still does not communicate after checking patient safety, network cable, wall jack, link status, room assignment, central station view, and known-good network connections, common external causes have been ruled out.
The device should be:
- Removed from service
- Labeled Out of Service
- Sent for repair, bench evaluation, or Spacelabs/network system review
Knowing when to stop is proper troubleshooting. A monitor that cannot reliably communicate with the central station should not be returned to patient care until communication and alarm visibility are verified.
Clinical Use Tip
Do not troubleshoot central station communication on an active patient unless clinical staff have confirmed safe alternate monitoring or direct observation. A bedside monitor may still show vitals locally while alarms are not visible at the central station.
Move the patient to a backup device or provide direct bedside observation before continuing any troubleshooting that could affect therapy continuity or alarm visibility.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Nursing reported that the Spacelabs Xprezzon in ICU Bed 12 was displaying patient vitals locally but was not appearing at the central station."
Cause
What was observed during troubleshooting.
Example:
"Found network cable loose at the wall jack with no active link light present."
Resolution
What action was taken.
Example:
"Reseated the network cable, verified link activity, confirmed waveform and alarm visibility at the central station, and returned the monitor to service."
Helpful Details to Include (If Known)
- Patient connected at time of failure
- Bedside monitor displaying normally
- Central station symptom observed
- Room and bed number
- Network cable condition
- Wall jack tested
- Link/activity lights present or absent
- Known-good cable used
- Known-good wall port tested
- Other monitors affected or not affected
- Recent monitor move or bed reassignment
- Central station alarm visibility confirmed
- IT or vendor escalation number
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
Network communication failures should be handled carefully because bedside monitoring and central station visibility are not the same thing. Start with patient safety, then verify the physical network path, monitor assignment, central station behavior, and whether the issue is isolated or widespread. Clear documentation helps separate device failure from network or configuration problems.
That is successful troubleshooting.