On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
This guide assists Clinical Engineering in troubleshooting a STERIS AMSCO 3085 SP surgical table that drifts unintentionally or cannot maintain a set position. Symptoms may include:
- Table slowly lowering, raising, or tilting without input
- Difficulty keeping lateral, trendelenburg, or height positions stable
- Slight continuous movement when controls are idle
These issues often stem from hydraulic or mechanical external factors before considering internal component failure.
Step-by-Step Troubleshooting
Verify Table is on a Level Surface
- Uneven floor can cause apparent drift.
- Ensures accurate assessment of hydraulic stability.
Check Power Connection
- Confirm main power cord is securely connected and outlet is functional.
- A table without stable power can intermittently lose hydraulic pressure.
Inspect Hand Control / Pendant
- Ensure buttons are not stuck or continuously pressed.
- Faulty pendant signals may mimic drift.
Examine Hydraulic Fluid Level and Leaks
- Look for visible leaks around cylinders or under the table.
- Low hydraulic fluid can prevent the table from maintaining position.
Check for Obstructions
- Ensure nothing is pressing on the table’s moving parts or rails.
- Objects or linens caught under the table can contribute to drift.
Test All Movements Slowly
- Observe if drift occurs in multiple axes (height, tilt, lateral).
- Note if drift correlates to a specific movement or actuator.
Listen for Unusual Sounds
- Hissing, bubbling, or knocking may indicate internal hydraulic issues.
- Confirms whether problem is hydraulic or mechanical.
If the Problem Persists
External checks completed: power, pendant, level, and leaks ruled out.
Likely causes: internal hydraulic valve wear, cylinder seal failure, or internal linkage issues.
- Remove table from service
- Label as Out of Service
- Send for bench evaluation or repair
Knowing when to stop prevents unnecessary risk and ensures patient safety.
Clinical Use Tip
- Never troubleshoot the table while a patient is on it.
- Move the patient to a stable, functioning table first.
- Only test movement when the area is clear and safe.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
“Table drifts downward when height is set, cannot hold position.”
Cause
What was observed during troubleshooting.
Example:
“Hydraulic fluid level adequate; no external leaks detected; hand control functioning normally; drift persists in all axes.”
Resolution
What action was taken.
Example:
“Table removed from service, labeled Out of Service, sent to vendor for hydraulic valve inspection and repair.”
Helpful Details to Include
- Floor surface: level?
- Pendant behavior: any stuck buttons?
- Observable leaks or stains
- Drift behavior: slow/jerky, which axes affected
- Sounds during movement
- Final table status (OOS, in-service elsewhere)
Final Thought
Patient safety is paramount. Using a logical, stepwise approach allows Clinical Engineering to isolate external factors before escalating. Proper documentation and timely escalation prevent risk, ensure repair efficiency, and support ongoing cost avoidance.
That is successful troubleshooting.