On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
This guide addresses the Alarm: Exhaust Rate Too Fast condition where chamber exhaust occurs faster than the expected cooling rate during a liquid cycle, which can impact proper cooling and compromise sterilization integrity.
Step-by-Step Troubleshooting
Ensure Patient Safety First
- Remove the sterilizer from active use if processing critical items
- Do not rely on loads from a cycle with unresolved exhaust rate alarms
- Ensure no questionable loads are released for patient use
Verify the Reported Alarm and Behavior
- Confirm Alarm: Exhaust Rate Too Fast is displayed
- Verify it occurs during the exhaust phase of a liquid cycle
- Observe chamber pressure dropping faster than expected
Check Cycle Selection and Configuration
- Ensure a liquid cycle is selected
- Confirm no incorrect cycle parameters were modified
- Verify standard cycle configuration is in use
Inspect Chamber Load
- Ensure liquid containers are appropriate and not overfilled
- Confirm proper spacing between items
- Verify no obstruction of drain or contact with chamber walls
Check Facility Steam Supply Stability
- Verify consistent steam pressure
- Check for recent facility steam interruptions or fluctuations
Listen and Observe Exhaust Operation
- Listen for abnormal or continuous venting
- Observe rapid pressure drop on display
- Compare behavior to expected gradual exhaust
Check for Obvious Valve or Drain Issues
- Inspect for visible steam discharge during exhaust
- Check for abnormal noise or continuous venting
- Look for signs of leakage at exhaust points
Rule Out Environmental or Installation Factors
- Ensure drain system is not excessively assisted
- Confirm no abnormal piping slope or vacuum effects
- Verify installation meets standard conditions
If the Problem Persists
If the alarm continues after all external checks:
- External causes have been ruled out
- Likely causes include solenoid valve malfunction (S40 or S03), steam trap malfunction, or control calibration issues
- Remove the device from service
- Label the unit Out of Service
- Send for qualified repair or calibration
Clinical Use Tip
Never troubleshoot on an active patient. Move the load or process to a backup sterilizer before addressing this issue.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Sterilizer displayed 'Exhaust Rate Too Fast' alarm during liquid cycle; chamber pressure dropped rapidly during exhaust phase."
Cause
What was observed during troubleshooting.
Example:
"External factors ruled out; exhaust occurring faster than expected indicating possible solenoid valve or steam trap malfunction."
Resolution
What action was taken.
Example:
"Unit removed from service, labeled Out of Service, and referred for repair and calibration."
Helpful Details to Include (If Known)
- Alarm behavior
- Accessories swapped
- Power behavior
- Environmental factors
- Indicator lights
- Final device status
Final Thought
Exhaust rate control is critical for proper liquid cycle cooling and sterilization. Always verify setup and environmental conditions first, then escalate when control components are suspected to maintain safety and reliability.
That is successful troubleshooting.