STERIS AMSCO 400 Series

Alarm: Over Sterilize Temperature - Sterilize temperature is above setpoint by more than prescribed amount

On this page

Asset Type

Sterilizer

Manufacturer

STERIS

Model

AMSCO 400 Series

What This Guide Helps With

This guide helps troubleshoot the “Over Sterilize Temperature” alarm, which occurs when chamber temperature exceeds allowable limits. This condition can lead to improper sterilization control and potential damage to instruments or loads.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Steam Supply Pressure

4. Inspect Steam Supply Piping

5. Check Chamber Steam Trap Function

6. Evaluate Solenoid Valve Operation

7. Consider Control Calibration Drift

If the Problem Persists

If external causes have been ruled out and the alarm continues:

Clinical Use Tip

Never troubleshoot a sterilizer during active patient use. Always ensure instruments are processed on a verified working unit before clinical use.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterilizer alarmed ‘Over Sterilize Temperature’ during sterilize phase; cycle continued but exceeded setpoint."

Cause

What was observed during troubleshooting.

Example:
"Steam supply pressure elevated above normal range / no external cause identified indicating possible internal fault."

Resolution

What action was taken.

Example:
"Verified steam supply and piping; unit removed from service and escalated for repair and calibration."

Helpful Details to Include (If Known)

Final Thought

Over-temperature alarms indicate a loss of controlled sterilization conditions. Always rule out external causes first, but escalate promptly if proper temperature control cannot be maintained.

That is successful troubleshooting.