STERIS AMSCO 400 Series

Alarm: Pressure In Chamber – 2 psig pressure sensed in chamber when not in cycle

On this page

Asset Type

Sterilizer

Manufacturer

STERIS

Model

AMSCO 400 Series

What This Guide Helps With

Alarm: Pressure In Chamber indicating approximately 2 psig present outside of a cycle, creating a potential safety hazard and preventing proper system readiness.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Chamber Pressure Reading vs Actual Condition

4. Check for Obvious External Causes

5. Observe Exhaust Function

6. Isolate Likely Cause

7. Stop When Resolved

If the Problem Persists

If the alarm continues or pressure repeatedly returns:

Remove device from service, label it Out of Service, and send for repair or qualified bench evaluation.

Clinical Use Tip

Never troubleshoot or open a sterilizer chamber under pressure. Always move workflow to a backup device and verify atmospheric pressure before access.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Pressure In Chamber alarm displayed while unit was idle; chamber indicated ~2 psig."

Cause

What was observed during troubleshooting.

Example:
"Persistent pressure reading outside cycle; suspected solenoid valve S02 or pressure transducer malfunction."

Resolution

What action was taken.

Example:
"Verified exhaust function and pressure reading. Alarm persisted. Removed unit from service and sent for repair."

Helpful Details to Include (If Known)

Final Thought

Pressure-related alarms present real safety risks. Always verify true chamber conditions, eliminate external causes first, and escalate when internal faults are suspected. Clear documentation ensures proper follow-up and accountability.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide