STERIS AMSCO 400 Series

Alarm: Recorder Deviation Alarm – Temperature sensors differ by more than 1°F

On this page

Asset Type

Sterilizer

Manufacturer

STERIS

Model

AMSCO 400 Series

What This Guide Helps With

This guide addresses the Recorder Deviation Alarm, which occurs when the two temperature sensing elements in the chamber drain probe differ by more than 1°F. This condition can lead to inaccurate temperature monitoring and compromise sterilization validation.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Recent Load and Environmental Conditions

Inspect Drain Area for Contamination

Check Accessible Wiring and Connections

Power Cycle the Sterilizer

Run a Test Cycle

If the Problem Persists

If the alarm continues after all external checks:

Clinical Use Tip

Never troubleshoot or rely on a sterilizer during active patient care. Always move instruments to a verified working sterilizer before continuing processing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Recorder Deviation Alarm displayed during sterilization cycle."

Cause

What was observed during troubleshooting.

Example:
"Temperature sensors in drain probe showing mismatch greater than 1°F with no external contributing factors."

Resolution

What action was taken.

Example:
"Removed unit from service and referred for probe evaluation and calibration."

Helpful Details to Include (If Known)

Final Thought

Accurate temperature sensing is critical for safe sterilization. When sensor disagreement occurs, always verify external factors first and escalate appropriately to ensure reliable operation and patient safety.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide