STERIS AMSCO 400 Series

Alarm: Too Long In Air Break - chamber does not air break vacuum to 2 inHg within allotted time

On this page

Asset Type

Sterilizer

Manufacturer

STERIS

Model

AMSCO 400 Series

What This Guide Helps With

Alarm: Too Long In Air Break causing failure to break chamber vacuum, delaying cycle completion and preventing proper pressure equalization.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Air Inlet Path (Most Common External Cause)

4. Verify Facility Air and Exhaust Conditions

5. Inspect for Obvious Air Leaks or Seal Issues

6. Isolate Toward Internal Components

If the Problem Persists

If the alarm continues after external checks, external causes have been ruled out and the issue likely indicates internal failure.

Clinical Use Tip

Never troubleshoot a sterilizer on an active load intended for patient use. Always move instruments to a backup sterilizer before troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterilizer alarmed 'Too Long In Air Break' and cycle extended without completing."

Cause

What was observed during troubleshooting.

Example:
"Air inlet filter found clogged restricting airflow; air break could not reach required pressure."

Resolution

What action was taken.

Example:
"Cleaned/replaced air inlet filter and verified airflow; unit tested successfully. If unresolved, sent for solenoid/control repair."

Helpful Details to Include (If Known)

Final Thought

Air break alarms often begin with simple airflow restrictions but can indicate deeper control or valve issues if persistent. Focus on external airflow first, then escalate appropriately to ensure safe sterilization performance.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide