STERIS AMSCO 400 Series

Alarm: Too Long In Charge - Chamber did not reach set temperature within allotted time

On this page

Asset Type

Sterilizer

Manufacturer

STERIS

Model

AMSCO 400 Series

What This Guide Helps With

This guide addresses the “Too Long In Charge” alarm, which occurs when the chamber does not reach the required sterilization temperature within the allotted time. This can result in delayed cycles or incomplete sterilization, posing a direct risk to patient safety if instruments are used.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Steam Supply Valve

4. Check Steam Pressure to Unit

5. Inspect Steam Supply Piping

6. Evaluate Steam Regulator Function

7. Check Solenoid Valve Operation

8. Rule Out Control or Calibration Issues

9. Confirm Resolution

If the Problem Persists

If steam supply, valves, and external conditions have been verified:

Clinical Use Tip

Never troubleshoot a sterilizer while it is actively being relied on for instrument processing. Always verify cycle completion and use a backup system before addressing alarms.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterilizer showing 'Too Long In Charge' alarm and not reaching temperature during cycle."

Cause

What was observed during troubleshooting.

Example:
"Steam supply valve partially closed and insufficient pressure reaching unit."

Resolution

What action was taken.

Example:
"Opened steam supply valve, verified proper pressure, and confirmed normal operation with test cycle."

Helpful Details to Include (If Known)

Final Thought

“Too Long In Charge” is a critical sterilization alarm that indicates the system cannot reach required conditions in time. Always start with steam supply and external factors before escalating to internal components. Proper troubleshooting ensures both device reliability and patient safety.

That is successful troubleshooting.

Related Guides

Was this guide helpful?

Don't see a guide you need?

Suggest a Guide