STERIS AMSCO 400 Series

Alarm: Too Long In Evacuation - Chamber did not reach required vacuum level within allotted time

On this page

Asset Type

Sterilizer

Manufacturer

STERIS

Model

AMSCO 400 Series

What This Guide Helps With

Alarm: Too Long In Evacuation indicating failure to reach required vacuum level, impacting sterilization cycle completion and proper air removal.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Water Supply

Inspect Chamber Drain Strainer

Check Door Seal Function

Inspect for External Leaks

Evaluate Valves (Non-Invasive Checks)

Attempt Cycle Again

If the Problem Persists

If alarm continues after external causes have been ruled out (water supply, drain, seal, leaks):

Clinical Use Tip

Never troubleshoot on an active patient. Always move instruments and workflow to a properly functioning sterilizer before continuing.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterilizer displaying 'Too Long In Evacuation' alarm and unable to complete cycle."

Cause

What was observed during troubleshooting.

Example:
"Chamber drain strainer partially clogged restricting evacuation flow."

Resolution

What action was taken.

Example:
"Cleaned drain strainer, verified water pressure, and confirmed normal operation on test cycle."

Helpful Details to Include (If Known)

Final Thought

Evacuation alarms directly impact sterilization effectiveness. Always verify water supply, drainage, and sealing before suspecting internal failures. Escalate appropriately when external causes are ruled out to maintain safety and reliability.

That is successful troubleshooting.