STERIS AMSCO 400 Series

Alarm: Too Long In Exhaust - chamber does not exhaust to 4 psig within the allotted time

On this page

Asset Type

Sterilizer

Manufacturer

STERIS

Model

AMSCO 400 Series

What This Guide Helps With

This guide addresses the “Too Long In Exhaust” alarm condition where the chamber fails to reach approximately 4 psig within the expected time, resulting in extended exhaust phases and potential incomplete sterilization cycles.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Alarm and Behavior

Check Chamber Drain Strainer (Most Common Cause)

Inspect Drain and Exhaust Path

Evaluate Load Conditions

Verify Door Gasket and Seal Condition

Attempt a Test Cycle

If the Problem Persists

If all external causes have been ruled out:

Clinical Use Tip

Never troubleshoot or release instruments from a sterilizer that alarmed during a cycle. Always move to a backup sterilizer before continuing clinical operations.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterilizer alarming 'Too Long In Exhaust'; chamber not reaching ~4 psig and extending exhaust phase."

Cause

What was observed during troubleshooting.

Example:
"Chamber drain strainer clogged, restricting exhaust flow and slowing depressurization."

Resolution

What action was taken.

Example:
"Cleaned drain strainer and verified exhaust path; test cycle completed successfully without alarm."

Helpful Details to Include (If Known)

Final Thought

This alarm indicates a failure to properly depressurize the chamber, which directly impacts sterilization effectiveness. Always rule out flow restrictions first, then escalate to internal component or calibration issues.

That is successful troubleshooting.