STERIS AMSCO 400 Series

Alarm: Under Sterilize Temperature - chamber temperature dropped below sterilize temperature

On this page

Asset Type

Sterilizer

Manufacturer

STERIS

Model

AMSCO 400 Series

What This Guide Helps With

This guide addresses the alarm condition where the chamber temperature drops below the required sterilization temperature during the cycle. This condition can result in incomplete sterilization and automatic cycle restart or extension, impacting patient safety and workflow.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Alarm and Behavior

3. Check Steam Supply (Most Common Cause)

4. Inspect Steam Supply Piping

5. Observe Chamber Behavior During Cycle

6. Check for Signs of Steam Trap Issues

7. Rule Out Facility or External Factors

8. Restore Service if Resolved

If the Problem Persists

If all external causes have been ruled out and the alarm continues, internal component issues such as steam trap failure, solenoid valve malfunction, or control calibration problems are likely.

Remove the device from service, label it Out of Service, and send it for qualified repair or calibration.

Clinical Use Tip

Never troubleshoot or rely on a sterilizer during an active clinical need. Always transfer sterilization needs to a backup unit before addressing the issue.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterilizer alarming ‘Under Sterilize Temperature’ during cycle with restart observed."

Cause

What was observed during troubleshooting.

Example:
"Steam pressure found below required level or suspected internal component issue after external checks."

Resolution

What action was taken.

Example:
"Verified steam supply and piping. Issue persisted. Removed unit from service and escalated for repair."

Helpful Details to Include (If Known)

Final Thought

Consistent sterilization temperature is critical for safe operation. Always prioritize verifying steam supply and external conditions first, and escalate promptly when internal faults are suspected.

That is successful troubleshooting.