STERIS AMSCO 4085

Error: DUAL COMMANDE

On this page

Asset Type

Surgical Table

Manufacturer

STERIS

Model

AMSCO 4085

What This Guide Helps With

This guide addresses the Error: DUAL COMMANDE condition, which occurs when both the foot pedal and hand control are used simultaneously, creating a control conflict where the system prioritizes hand control input.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check for Simultaneous Control Use

Inspect Foot Pedal for Stuck or Continuous Input

Inspect Hand Control Buttons

Test Controls Independently

Stop When Resolved

If the Problem Persists

If the error continues even when only one control is used, external causes have been ruled out. This may indicate an internal control or input circuit issue.

Clinical Use Tip

Never troubleshoot table controls while a patient is in a vulnerable surgical position. Move the patient to a backup device before performing troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Table displaying DUAL COMMANDE error during positioning; inconsistent response from controls."

Cause

What was observed during troubleshooting.

Example:
"Simultaneous input detected from foot pedal and hand control; foot pedal found partially engaged causing continuous signal."

Resolution

What action was taken.

Example:
"Freed and verified proper function of foot pedal; confirmed normal operation using single control input; returned to service."

Helpful Details to Include (If Known)

Final Thought

This error is often the result of normal system logic rather than a true failure. Understanding how control inputs interact can prevent unnecessary service calls and improve safe device operation.

That is successful troubleshooting.