STERIS AMSCO 4085

Error: LOCK AGAIN

On this page

Asset Type

Surgical Table

Manufacturer

STERIS

Model

AMSCO 4085

What This Guide Helps With

Error: LOCK AGAIN indicating floor lock failure due to pressure switch not changing state, preventing proper table stabilization or locking.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Check for External Hydraulic Oil Leakage

Inspect Floor Lock Operation Physically

Check Pressure Switch Wiring and Connections

Rule Out Control/Input Issues

Stop if the Issue is Resolved

If the Problem Persists

If external causes such as oil leaks, wiring issues, and control inputs have been ruled out:

Clinical Use Tip

Never troubleshoot or cycle the floor lock while a patient is on the table. Always transfer the patient to a backup device before performing checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Table displaying 'LOCK AGAIN' error. Floor lock not engaging."

Cause

What was observed during troubleshooting.

Example:
"No external leaks or wiring issues found. Pressure switch not changing state, indicating likely valve block or sensor failure."

Resolution

What action was taken.

Example:
"Removed table from service, labeled out of service, and sent for repair."

Helpful Details to Include (If Known)

Final Thought

Floor lock errors are critical safety issues that must be addressed immediately. Always begin with visible hydraulic and wiring checks, but recognize when internal failures require escalation to ensure safe operation.

That is successful troubleshooting.