What This Guide Helps With

Error: MOTOR FAIL causing continuous movement or uncontrolled motor operation due to a stuck motor power condition. This can result in unintended table movement and presents a serious safety hazard.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Check Hand Control and Foot Pedal for Stuck Inputs

4. Inspect Cables and Connections

5. Power Cycle the Table

6. Isolate Function (If Applicable)

7. Evaluate for Internal Control Failure

If the Problem Persists

If continuous motor operation with Error: MOTOR FAIL persists after all external checks:

Clinical Use Tip

Never troubleshoot uncontrolled table movement while a patient is on the table. Always transfer the patient to a safe surface before attempting any troubleshooting.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Table displaying Error: MOTOR FAIL with continuous movement without user input."

Cause

What was observed during troubleshooting.

Example:
"Internal motor control fault. External controls and cables verified functional. Likely sliding board failure."

Resolution

What action was taken.

Example:
"Removed table from service, tagged out of service, and sent for repair. Sliding board replacement required."

Helpful Details to Include (If Known)

Final Thought

Errors like MOTOR FAIL represent serious control faults and must be treated as safety-critical conditions. Always rule out external causes first, but escalate quickly when uncontrolled movement persists to prevent harm.