STERIS AMSCO 4085

Error: PUMP FAIL

On this page

Asset Type

Surgical Table

Manufacturer

STERIS

Model

AMSCO 4085

What This Guide Helps With

Error: PUMP FAIL where the hydraulic pump runs continuously, causing unintended table movement or potential system overheating.

Step-by-Step Troubleshooting

1. Ensure Patient Safety First

2. Verify the Reported Error and Behavior

3. Check Hand Control and Foot Pedal for Stuck Inputs

4. Inspect Control Cables and Connections

5. Power Cycle the Table

6. Verify Power Supply Behavior

If the Problem Persists

If all external controls, cables, and power conditions have been ruled out and the pump continues running:

Clinical Use Tip

Never troubleshoot a surgical table on an active patient. Uncontrolled hydraulic movement can create serious safety risks. Always transfer the patient to a backup device first.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Table displaying Error: PUMP FAIL with hydraulic pump running continuously without input."

Cause

What was observed during troubleshooting.

Example:
"Pump continued running after verifying controls and cables, indicating internal control circuit failure."

Resolution

What action was taken.

Example:
"Removed from service and sent for repair. Power supply replaced to resolve issue."

Helpful Details to Include (If Known)

Final Thought

Continuous pump operation is a high-risk condition. Always rule out external inputs first, but recognize when persistent behavior indicates internal failure requiring escalation.

That is successful troubleshooting.