Error: WRONG USE

On this page

Asset Type

Surgical Table

Manufacturer

STERIS

Model

AMSCO 4085

What This Guide Helps With

Error: WRONG USE displayed when a control button is held too long, causing table movement lockout or unresponsive controls.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Verify the Reported Error and Behavior

Release All Control Buttons

Inspect Hand Control and Foot Pedal

Check Cable and Connection Integrity

Test Normal Operation

If the Problem Persists

If the error continues without any buttons being pressed, external causes have been ruled out. The issue may indicate a faulty hand control, foot pedal, or control interface circuit.

Remove the device from service, label it Out of Service, and send for repair or bench evaluation.

Clinical Use Tip

Never troubleshoot on an active patient. Move the patient to a backup device before performing any checks.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Table displaying 'WRONG USE' error and not responding to controls."

Cause

What was observed during troubleshooting.

Example:
"Continuous button activation detected due to prolonged press or stuck control button."

Resolution

What action was taken.

Example:
"Released controls, inspected hand control and cables, verified proper function, and returned table to service."

Helpful Details to Include (If Known)

Final Thought

Errors like “WRONG USE” are designed to protect both the equipment and patient. Always verify user input and external controls before suspecting device failure.

That is successful troubleshooting.