On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Use this guide when an approved vacuum leak test fails because of door sealing, chamber condition, drainage, moisture, or service-level vacuum faults.
Step-by-Step Troubleshooting
1. Protect Patient Safety and Sterilization Reliability
A failed vacuum leak test indicates the sterilizer may not reliably maintain the chamber conditions required for applicable cycles.
Do not use the unit for production loads requiring verified air removal until the failure is resolved according to facility policy.
Expected outcome: Questionable sterilization performance is prevented from affecting patient care.
2. Confirm the Exact Test Failure
Review the test result and cycle record. Determine whether:
- The correct test cycle was selected.
- The chamber was empty as required by facility procedure.
- The test failed consistently or only once.
- A specific alarm or message accompanied the failure.
- The sterilizer had recently completed a hot or wet cycle.
Do not invent or substitute acceptance limits; use the applicable manufacturer and facility criteria.
Expected outcome: A genuine vacuum leak test failure is confirmed under appropriate test conditions.
3. Verify Chamber Is Empty and Properly Prepared
Remove any load, packaging, loose materials, or accessories not intended to remain in the chamber during the test.
Check that the chamber is in the appropriate condition for the approved leak-test procedure.
Expected outcome: No load-related item can interfere with sealing or test conditions. If preparation error caused the result, repeat the test correctly and stop troubleshooting if it passes.
4. Inspect Door and Accessible Sealing Surfaces
With the sterilizer safe, inspect the chamber opening and accessible sealing surfaces for:
- Debris.
- Packaging fragments.
- Tape or labels.
- Residue.
- Visible damage.
- Objects preventing complete closure.
Do not remove or adjust internal door mechanisms.
Expected outcome: The door closes fully against clean, unobstructed sealing surfaces. If cleaning or removing debris restores a passing test, complete required verification and stop troubleshooting.
5. Verify Door Closure and Lock Recognition
Operate the door normally and confirm the sterilizer recognizes the expected closed and locked condition.
Do not bypass door sensing or locking interlocks.
Expected outcome: The door locks normally and the control system consistently recognizes the correct state.
6. Inspect Chamber and Drain for Residual Water
Look for abnormal standing water or excessive moisture in the chamber and inspect the accessible drain area for debris.
Residual water or poor drainage may interfere with repeatable vacuum testing.
Expected outcome: The chamber and drain appear normal with no obvious blockage or abnormal water accumulation. If an accessible obstruction is corrected, repeat the approved test.
7. Check for Obvious External Leakage
Observe accessible door, drain, and utility areas for abnormal air, steam, or water leakage that may indicate a sealing problem.
Do not disassemble vacuum piping or internal valves.
Expected outcome: No obvious external leak is present. Any significant leak or damaged chamber/door hardware requires removal from service.
8. Allow Appropriate Stabilization Before Retesting
If the previous cycle or chamber condition may have affected the test, follow facility and manufacturer procedures regarding preparation before repeating the test.
Do not repeatedly run tests without addressing an identified cause.
Expected outcome: The leak test is repeated under appropriate, consistent conditions.
9. Repeat the Approved Vacuum Leak Test
Run the proper vacuum leak test using established facility and manufacturer criteria. Record the actual result.
Expected outcome: The unit passes the approved leak test. If it passes and no other abnormal condition exists, troubleshooting can stop after all required return-to-service checks.
10. Escalate Repeated Leak-Test Failure
If chamber preparation, door closure, sealing surfaces, drain condition, and external leakage have been checked and the approved test continues to fail, stop external troubleshooting.
Potential service-level categories include internal valves, chamber plumbing, vacuum components, door sealing systems, drains, sensing, or control functions.
Expected outcome: The sterilizer remains out of production until the vacuum system is properly evaluated.
If the Problem Persists
Common external causes have been ruled out. Remove the sterilizer from service, label it Out of Service, and arrange qualified service or repair evaluation.
Further diagnostics should use current STERIS service documentation and approved test equipment. Internal vacuum valves, piping, pumps or ejector systems, door systems, drains, and sensors should be evaluated only by qualified personnel.
After repair, repeat all manufacturer- and facility-required leak and sterilization-performance tests before return to service. Knowing when to stop repeated testing and escalate is proper troubleshooting.
Clinical Use Tip
Do not substitute a successful production cycle for a required vacuum leak test when the sterilizer has already demonstrated a leak-test failure.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Sterile Processing reported that the AMSCO 600 Series repeatedly failed its vacuum leak test."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found a strip of packaging material on the chamber door sealing surface that prevented complete sealing."
Resolution
What action was taken.
Example:
"Removed the debris, cleaned the accessible sealing surface, repeated the approved vacuum leak test successfully, and verified the sterilizer was ready for service."
Helpful Details to Include (If Known)
- Test cycle selected.
- Exact displayed result or message.
- Chamber condition before testing.
- Door and sealing-surface findings.
- Drain condition.
- Visible moisture or standing water.
- Number of failed tests.
- Correction performed.
- Final approved test result.
- Final device status.
Final Thought
A leak-test failure deserves a systematic approach: protect production loads, confirm correct test conditions, inspect the chamber and door externally, retest appropriately, and escalate persistent failures rather than guessing at internal components.
That is successful troubleshooting.