STERIS AMSCO Century V Series

Drain failure, water retention, or chamber flooding

On this page

Asset Type

Sterilizer

Manufacturer

STERIS

Model

AMSCO Century V Series

What This Guide Helps With

Identifies causes of slow or failed drainage, water retention, or chamber flooding that prevent cycle completion and compromise sterilization safety.

Step-by-Step Troubleshooting

1. Ensure Patient and Staff Safety

2. Inspect Chamber Drain Strainer and Trap

3. Verify Drain Line and Floor Connection

4. Check Solenoid Valve Operation

5. Inspect Water Level Sensors and Float Switches

6. Examine Inlet Water Supply

7. Run a Test Cycle

If the Problem Persists

External checks have been completed. Likely internal pump, valve, or controller issue.

Clinical Use Tip

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Chamber retains water after cycle, causing flooding."

Cause

What was observed during troubleshooting.

Example:
"Drain strainer partially blocked with lint and debris, valve actuation delayed."

Resolution

What action was taken.

Example:
"Cleared strainer and verified drain line; test cycle completed with normal drainage."

Helpful Details to Include (If Known)

Final Thought

Proper troubleshooting begins with external checks to prevent escalation. Patient safety and logical investigation are paramount. Accurate documentation ensures traceability and supports future maintenance.

That is successful troubleshooting.