STERIS Reliance 444

Chamber Will Not Fill or Water Level Is Incorrect

On this page

Asset Type

Washer-Disinfector

Manufacturer

STERIS

Model

Reliance 444

What This Guide Helps With

Troubleshoots absent, slow, excessive, or inconsistent chamber filling caused by water supply, loading, drainage, or other external conditions.

Step-by-Step Troubleshooting

1. Protect Processing and Confirm the Fill Problem

Do not use the washer for instrument processing when chamber filling is unreliable. Identify whether the chamber:

Record the cycle and phase where the condition occurs.

Expected outcome: The exact fill complaint is confirmed. If the observation was related to normal cycle behavior and a complete verified cycle operates normally, troubleshooting can stop.

2. Verify Normal Power and Cycle Initiation

Confirm the washer remains powered and is actually progressing into a fill phase rather than being held by another condition such as an open door or incomplete rack connection.

Expected outcome: The unit is stable and is commanding normal cycle progression. If correcting an external start condition restores filling, verify the cycle and stop.

3. Verify Facility Water Availability

Check accessible incoming water services for:

Coordinate with Facilities if supply pressure or infrastructure is suspected.

Expected outcome: Required facility water supplies are available. If an external supply condition is corrected and normal fill is restored, complete a verification cycle and stop.

4. Inspect Accessible Supply Connections

Inspect visible hoses, fittings, strainers or accessible connection areas without disassembly for kinks, crushing, leakage, or obstruction.

Do not open pressurized plumbing or internal valves unless qualified and authorized.

Expected outcome: External supply connections are intact and unobstructed. Correcting a visible external restriction should restore normal filling.

5. Check the Chamber and Load

Inspect the chamber for misplaced items, packaging, containers, or debris that could interfere with visible level sensing areas, drains, or water distribution.

Verify the rack and load are correctly positioned.

Expected outcome: Nothing in the chamber or load is interfering with normal filling or drainage. If removing an obstruction corrects the water level, verify operation and stop.

6. Check for Simultaneous Drainage

Observe whether incoming water appears to leave the chamber unusually quickly. A drain condition can appear to be a fill problem.

Inspect accessible drain areas for obvious blockage or abnormal flow.

Expected outcome: The chamber retains water appropriately during the required portion of the cycle. If a simple accessible drain obstruction caused the symptom and is safely corrected, verify the cycle and stop.

7. Compare Repeated Cycle Behavior

Run an approved verification cycle and determine whether the fill problem occurs:

Expected outcome: The failure pattern is defined. If repeated cycles fill consistently after correcting an external cause, troubleshooting is complete.

8. Perform Final Functional Verification

Confirm the chamber fills and drains normally throughout the verification cycle without leakage, overfill, or interruption.

Expected outcome: Normal cycle progression is restored and the washer completes its required verification. The unit may be returned to service only after successful testing.

9. Escalate Unresolved Fill Problems

If water supply, visible connections, loading, and drainage conditions are normal but fill remains incorrect, stop external troubleshooting.

Expected outcome: The washer is removed from service for qualified evaluation.

If the Problem Persists

The remaining issue may involve internal water-level sensing, valves, plumbing, controls, utility regulation, or other service-level components.

The washer should be:

Return-to-service testing must demonstrate correct filling, draining, cycle progression, and absence of leaks. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Do not process instruments in a washer with unreliable chamber water levels because cleaning performance cannot be assumed.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterile Processing reported the Reliance 444 chamber was not filling adequately during the wash cycle."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the facility water isolation valve supplying the washer was partially closed."

Resolution

What action was taken.

Example:
"Facilities restored the valve to its proper position, and Clinical Engineering verified normal chamber filling and a complete successful cycle."

Helpful Details to Include (If Known)

Final Thought

Verify the water supply, visible plumbing, chamber condition, loading, and drainage before assuming an internal water-control problem. Remove the washer from service when water level cannot be reliably verified and document the completed troubleshooting clearly.

That is successful troubleshooting.