STERIS Reliance 444

Cycle Will Not Start, Aborts, or Stalls

On this page

Asset Type

Washer-Disinfector

Manufacturer

STERIS

Model

Reliance 444

What This Guide Helps With

Troubleshoots cycles that will not begin, stop unexpectedly, or remain in one phase because of power, doors, utilities, loading, or external conditions.

Step-by-Step Troubleshooting

1. Protect Workflow and Confirm the Reported Condition

Do not rely on a washer-disinfector that cannot complete a validated cycle. Remove affected instruments from the normal processed workflow and follow facility procedures for alternative decontamination processing.

Confirm whether the cycle:

Record the selected cycle and the point where operation stops.

Expected outcome: The failure mode and affected cycle phase are clearly identified. If the problem was an incorrect cycle selection or incomplete start sequence and the unit subsequently operates normally, verify a complete cycle and stop troubleshooting.

2. Verify Power and Normal Startup

Confirm the washer is powered, the display and controls respond normally, and no facility breaker, disconnect, or emergency power condition is affecting operation.

Look for:

Do not repeatedly reset a breaker or continue operating equipment with electrical damage.

Expected outcome: Stable power and normal control operation are present. If restoring an external power condition corrects the problem and a complete verification cycle passes, troubleshooting can stop.

3. Check Door Position and Interlocks

Verify all required doors are fully closed and unobstructed. Inspect visible door edges, seals, latching surfaces, and the load for anything preventing full closure.

Do not bypass or defeat a door interlock.

Expected outcome: The door closes normally and the system recognizes the required door state. If correcting an obstruction restores normal cycle operation, complete functional verification and stop.

4. Verify Rack, Cart, and Load Position

Confirm the rack or processing cart is fully seated and properly aligned. Check that instruments, containers, hoses, or accessories are not interfering with chamber components, doors, spray arms, or manifold connections.

Expected outcome: The load is correctly positioned without mechanical interference. If reseating or correcting the load allows the cycle to complete normally, stop after verification.

5. Check Available Utilities

Verify the washer has the facility utilities required for operation. Inspect accessible water, steam, air, drain, and other applicable service connections for obvious shutoff, leakage, disconnection, or facility outage conditions.

Coordinate with Facilities when a utility issue is suspected.

Expected outcome: Required utilities are available without obvious interruption. If an external utility issue is corrected and the washer completes a test cycle, troubleshooting can stop.

6. Check Chemistry Availability

Confirm required detergent or other processing chemistry is present, correctly connected, and not visibly empty, kinked, disconnected, or leaking.

Do not substitute unapproved chemistry.

Expected outcome: Chemistry supply is available and externally connected correctly. If restoring the chemistry supply allows normal operation, verify the completed cycle and stop.

7. Compare Cycle Behavior

If permitted by facility procedure, perform an unloaded or approved test cycle and observe whether the unit progresses normally.

Determine whether the problem:

Expected outcome: The failure is isolated to the washer, a cycle configuration, or an external loading condition. A successful repeat cycle after correcting an external cause resolves the issue.

8. Perform Final Functional Verification

After any correction, run the appropriate verification cycle according to facility and manufacturer requirements. Confirm normal filling, washing, draining, temperature progression, door operation, chemistry delivery, and cycle completion as applicable.

Expected outcome: The cycle completes without alarms or abnormal interruption and required cycle documentation is produced. The washer may be returned to service only when all required checks pass.

9. Escalate if the Cycle Still Aborts or Stalls

If power, doors, loading, rack position, utilities, and chemistry have been ruled out, stop external troubleshooting.

Expected outcome: An unresolved washer is removed from service rather than repeatedly cycled or returned to clinical processing.

If the Problem Persists

Common external causes have been ruled out. The remaining issue may involve internal sensing, control, utility regulation, valve operation, safety interlocks, configuration, or another service-level condition.

The washer should be:

Do not bypass safety devices or enter unauthorized service configurations. After repair, perform all required operational and performance verification before returning the washer to service. Knowing when to stop external troubleshooting is proper troubleshooting.

Clinical Use Tip

Do not release a load from a cycle that aborted or stalled unless it has been reprocessed according to facility infection-prevention procedures.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Sterile Processing reported the Reliance 444 started a cycle but repeatedly stopped before completion."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the processing rack was not fully seated and was interfering with normal cycle initiation."

Resolution

What action was taken.

Example:
"Rack position was corrected, the washer completed a full verification cycle without interruption, and the unit was returned to service."

Helpful Details to Include (If Known)

Final Thought

Protect the processing workflow first, verify power, doors, loading, utilities, and chemistry before assuming an internal failure, and escalate when the washer cannot complete a verified cycle. Clear CCR documentation should show what was reported, what was found, and how safe operation was confirmed.

That is successful troubleshooting.