On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Troubleshooting failed barcode scans or load-ID entry caused by labels, scanner connection, contamination, focus, required fields, or interface problems.
Step-by-Step Troubleshooting
1. Protect Load Traceability
Do not process or release a load without the identification required by facility policy.
Notify Sterile Processing staff of the entry failure.
Use the approved manual downtime process when available.
Keep each load uniquely identified.
Do not substitute another load number merely to start the cycle.
Redirect work if accurate traceability cannot be maintained.
Expected outcome: Every load remains uniquely traceable despite the scanner or entry problem.
2. Confirm the Exact Entry Failure
Determine whether:
- The scanner has no light or aiming beam.
- The scanner reads but enters no data.
- Incorrect characters appear.
- The barcode is rejected.
- Manual touchscreen entry also fails.
- Only one label type is affected.
- The field accepts data but does not allow the cycle to proceed.
Record the exact message and affected field.
Expected outcome: The issue is separated into scanner power, barcode quality, data entry, field validation, or software workflow.
3. Inspect the Barcode Label
Check the label before testing hardware.
Confirm the barcode is clean, dry, flat, and undamaged.
Look for wrinkles, glare, low contrast, truncated printing, or marks across the code.
Verify the complete barcode is visible.
Try another known-good label of the same approved type.
Do not create false load identifiers for clinical processing.
Expected outcome: A valid, readable barcode is available. If another label scans normally, replace the defective label and troubleshooting can stop after confirming the correct record.
4. Check Scanner Position and Technique
Verify normal scanning distance and orientation.
Aim at the complete barcode.
Reduce glare from overhead lighting or glossy labels.
Hold the scanner steady.
Try a slight angle rather than scanning perpendicular to a reflective surface.
Confirm the aiming beam crosses the entire code.
Expected outcome: The scanner reads an approved known-good barcode consistently.
5. Inspect and Clean the Scanner Window
Look for fingerprints, adhesive, dust, scratches, or moisture.
Clean the window using an approved method.
Dry it fully.
Inspect the scanner housing and cable for damage.
Do not spray cleaner directly into the scanner.
Expected outcome: The scan window is clear and undamaged. If cleaning restores reliable scans, troubleshooting can stop after repeated verification.
6. Check the Scanner Connection
For a wired scanner:
Confirm the cable is fully seated at the scanner and sterilizer.
Inspect connectors and strain relief for damage.
Reseat the connection while the sterilizer is idle.
Verify the cable is not pinched or under tension.
Do not connect an unapproved consumer scanner.
For a wireless scanner, verify approved charging, pairing, and connection status without changing protected settings.
Expected outcome: The scanner powers normally and communicates with the sterilizer.
7. Test a Known-Good Approved Scanner
When an equivalent approved scanner is available:
Connect it using the same sterilizer port.
Scan a known-good approved barcode.
Test the suspect scanner on another compatible unit when permitted.
Keep scanner and cable substitutions clearly identified.
Restore the correct assigned accessories after testing.
Expected outcome: The problem follows either the scanner or the sterilizer connection. If the scanner is defective, replace it with an approved unit and verify operation.
8. Test Manual Load-ID Entry
Use the operator-accessible keyboard or touchscreen.
Select the correct field.
Enter a valid test or downtime identifier according to facility procedure.
Check whether characters appear correctly.
Confirm the entry can be accepted and saved.
Do not use a real load identifier for a nonclinical test.
Expected outcome: Manual entry works, indicating the control interface and field are functional. If neither scanner nor manual entry works, escalate the control or software issue.
9. Verify Field Requirements
Check for data-format problems.
Confirm the correct field is active.
Check whether the entry contains unsupported spaces or characters.
Verify required operator, load, or tracking fields are complete.
Compare the format with a known successful load record.
Do not change validation rules without authorization.
Expected outcome: The load ID meets facility and interface requirements and is accepted.
10. Check for External Interface Dependence
Determine whether the scanner or load ID depends on an instrument tracking system.
Confirm the tracking application is online.
Check whether other sterilizers can retrieve or validate load IDs.
Review network status.
Use approved downtime procedures if the server is unavailable.
Coordinate with Information Services and the application owner.
Expected outcome: A tracking-system outage is identified or ruled out.
11. Perform a Controlled Restart When Safe
If the scanner was recently connected or the entry field is frozen:
Confirm no cycle is active.
Use the normal shutdown process.
Restart the sterilizer with the approved scanner connected.
Observe whether the scanner initializes.
Test a known-good barcode and manual entry.
Expected outcome: Barcode and manual entry functions return and remain stable.
12. Complete Final Verification
Before normal use:
Scan multiple approved test barcodes.
Confirm the correct characters appear.
Verify the load ID is saved to the correct cycle record.
Confirm the record transfers to the tracking system when applicable.
Verify no duplicate, truncated, or altered identifier is created.
Expected outcome: Accurate load identification is maintained from entry through final cycle documentation.
If the Problem Persists
If label quality, scanner technique, cleaning, connections, known-good substitution, manual entry, required fields, and external tracking availability have been checked, common external causes have been ruled out.
The remaining cause may involve the sterilizer scanner port, input interface, control software, barcode configuration, data mapping, or tracking-system integration.
The affected scanner or sterilizer should be:
Removed from normal scanner-dependent workflow.
Labeled Out of Service when traceability cannot otherwise be maintained.
Sent for repair or qualified evaluation.
Evaluated using appropriate STERIS and scanner documentation.
Configured or repaired only by qualified personnel.
The sterilizer may remain in limited use only under an approved manual documentation process that preserves complete traceability. Verify entry, record storage, and transmission before restoring normal use.
Knowing when to stop external troubleshooting is proper troubleshooting.
Clinical Use Tip
Never reuse or invent a load ID to bypass an entry problem; each sterilization cycle must remain uniquely traceable.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"Sterile Processing reported that the V-PRO barcode scanner illuminated but would not enter the load ID."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found adhesive residue covering the scanner window and reducing barcode readability."
Resolution
What action was taken.
Example:
"The scanner window was cleaned, multiple approved labels were read correctly, and the load IDs were verified in the final cycle records."
Helpful Details to Include (If Known)
- Scanner light or aiming beam
- Exact entry message
- Barcode type and condition
- Labels tested
- Scanner window condition
- Cable and connector condition
- Known-good scanner result
- Manual entry result
- Required field format
- Tracking-system status
- Record and transmission verification
- Final service status
Final Thought
Protect traceability first, then check label quality, technique, scanner cleanliness, connections, manual entry, and tracking-system availability. Escalate persistent input failures and document the exact effect on the final cycle record.
That is successful troubleshooting.