Stryker 1688 Camera System

Image artifact or quality loss

On this page

Asset Type

Endoscopic / OR Camera System

Manufacturer

Stryker

Model

1688 Camera System

What This Guide Helps With

Identifies causes when video displays distortions, flickering, lines, or overall degraded image quality, affecting procedural visualization.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Inspect Camera Head and Lens

Verify Cable Connections

Test with Alternate Camera Head

Check Light Source and Fiber Optics

Adjust Console Settings

Power Cycle the System

Test on Different Monitor or Display

If the Problem Persists

External causes have been ruled out. The issue may be internal (camera sensor, processor board). Remove from service, label Out of Service, and send for repair. Avoid further use to protect patient safety.

Clinical Use Tip

Never troubleshoot on an active patient. Always ensure a functional backup system is available to maintain procedural safety.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"OR staff reported flickering lines and distorted colors during procedures."

Cause

What was observed during troubleshooting.

Example:
"Lens contamination and a frayed light cable were found during inspection."

Resolution

What action was taken.

Example:
"Lens cleaned and light cable replaced; image quality restored and system returned to service."

Helpful Details to Include

Final Thought

Prioritize patient safety and follow a logical troubleshooting sequence from simple to complex. Document all findings clearly. Escalate to repair only when external causes are eliminated.

That is successful troubleshooting.