On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Missing or unstable high-resolution video caused by cabling, display capability, routing equipment, input selection, configuration, or communication problems.
Step-by-Step Troubleshooting
1. Protect the Patient and Maintain Visualization
If video becomes unstable during a procedure, notify the clinical team and move visualization to a verified stable display path before troubleshooting. Do not troubleshoot flickering or intermittent surgical video while the team depends on it.
Expected outcome: A stable clinical image is available before technical testing begins.
2. Confirm the Exact Video Problem
Determine whether the output is:
- Completely absent
- Intermittently black
- Flickering
- Present at an unexpected resolution
- Stable on one display but not another
Identify which video output and display path are affected.
Expected outcome: The affected path is clearly identified.
3. Verify Camera Control Unit and Display Operation
Confirm the 1788 camera control unit has completed startup normally and produces a stable image on at least one known functional output when possible.
Verify the destination monitor is powered, set to the correct input, and otherwise functioning normally.
Expected outcome: Both source and display are operational independently of the suspect signal path.
4. Inspect the Video Cable
Inspect the entire accessible video cable for:
- Loose connectors
- Bent or damaged connector shells
- Crushed sections
- Sharp bends
- Damaged strain relief
- Excessive mechanical tension
Reseat both ends securely.
Expected outcome: Video remains stable after the cable is reseated. If the correct image returns and remains stable, troubleshooting can stop after final verification.
5. Bypass Intermediate Routing Equipment
If the video passes through a switcher, wall interface, matrix, extender, recorder, integration system, or other routing device, bypass intermediate equipment where practical and connect the camera system directly to a compatible known-good display.
Expected outcome: Direct connection produces stable expected video. If direct video works but the routed path does not, escalate the infrastructure or integration path rather than the camera control unit.
6. Substitute a Known-Good Video Cable
Replace the suspect video cable with a known-good approved cable appropriate for the connection.
Expected outcome: Stable video returns. If the issue follows the original cable, replace or remove that cable from service and troubleshooting can stop after verification.
7. Compare With a Known-Good Display
Connect the output to another verified compatible display when available.
If the same display receives other sources correctly but fails specifically with this connection, continue isolating the path.
Expected outcome: The test determines whether the problem follows the source, cable path, or destination display.
8. Verify User-Accessible Video Configuration
Review normal user-accessible output settings and confirm they are appropriate for the connected display and routing system.
Do not change restricted service settings or force unsupported output formats.
Expected outcome: The source and display operate with a mutually supported configuration.
9. Check for Environmental or Routing Interference
Look for loose wall plates, stressed connectors, adapters, unsupported conversion devices, or recently changed OR integration hardware.
Compare the affected room with a known-working room when practical.
Expected outcome: External infrastructure issues are identified or ruled out.
10. Perform Final Functional Verification
Verify:
- Continuous video
- Correct display path
- Stable expected image quality
- No flicker or intermittent black frames
- No unwanted fallback during normal use
- Normal operation through any required routing equipment
Expected outcome: Video remains stable through the complete intended clinical path. Troubleshooting can stop and the system may be returned to service according to facility policy.
If the Problem Persists
If cables, displays, routing equipment, input selection, and accessible configuration have been verified, the remaining cause may involve the camera control unit's video interface, display compatibility, OR integration infrastructure, or another service-level video-processing problem.
The affected equipment should be:
- Removed from service if the clinical video path is unreliable
- Labeled Out of Service
- Sent for repair or bench evaluation as appropriate
- Evaluated using appropriate manufacturer documentation and approved test equipment
- Repaired or configured only by qualified personnel
Coordinate with OR integration or audiovisual support when the failure is isolated outside the camera system. Complete full-path video verification before return to service.
Knowing when to stop external troubleshooting prevents an intermittent video path from being mistaken for a reliable repair.
Clinical Use Tip
Verify the complete camera-to-display path; a perfect image at the camera control unit does not prove the OR routing system is reliable.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported that the Stryker 1788 video intermittently flickered and dropped to a black screen on the primary surgical monitor."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found stable direct video from the camera system but reproduced the failure through the existing external video cable."
Resolution
What action was taken.
Example:
"Replaced the defective video cable and verified stable video through the complete clinical display path before returning the system to service."
Helpful Details to Include (If Known)
- Affected video output
- Destination display
- Whether direct connection worked
- Cable and connector condition
- Known-good cable results
- Known-good monitor results
- Intermediate switchers or routing equipment
- Whether other sources work on the same display path
- Behavior before and after correction
- Final video status
Final Thought
Start with the signal path rather than assuming the camera control unit has failed. Stable surgical video depends on every cable, interface, display, and routing device in the chain, and each should be verified before escalation.
That is successful troubleshooting.