Stryker 1788 Platform

AIM Fluorescence Mode Will Not Activate or Has Uneven Illumination

On this page

Asset Type

Endoscopic / OR Camera System

Manufacturer

Stryker

Model

1788 Platform

What This Guide Helps With

AIM fluorescence activation or illumination problems caused by setup, compatible accessories, connections, light delivery, contamination, positioning, or configuration.

Step-by-Step Troubleshooting

1. Protect the Patient and Maintain Visualization

If AIM fluorescence is required during an active procedure and is unavailable or unreliable, notify the clinical team. Do not repeatedly manipulate the system while fluorescence guidance is clinically required. Use a verified backup system or approved alternate workflow before troubleshooting.

Expected outcome: The clinical team has reliable visualization appropriate for the procedure.

2. Confirm the Exact AIM Failure

Determine whether AIM mode:

Document any displayed message and whether normal visible-light imaging remains functional.

Expected outcome: The failure is clearly defined and can be evaluated without patient dependence.

3. Verify the Complete Compatible Imaging Chain

Confirm that the connected camera head, camera control unit, light source, light cable, scope, and other required accessories are intended for the requested fluorescence workflow.

Do not assume AIM functionality simply because normal white-light video is present.

Expected outcome: The system is assembled with the appropriate compatible components. If correcting an accessory mismatch restores AIM operation, the issue is resolved and troubleshooting can stop after verification.

4. Check Camera and Light-Source Communication

Verify the camera head and light source are recognized normally. Inspect all external communication and video connections associated with the system.

Reseat accessible connections that can be safely disconnected and reconnect them fully.

Expected outcome: The camera system and light source remain connected and communicate without intermittent messages.

5. Inspect the Light Cable and Optical Connections

Check the light cable for:

Inspect the scope connection and other accessible optical interfaces according to approved cleaning practices.

Expected outcome: Light delivery is uniform and connections are secure. If cleaning or correctly seating an optical connection restores even illumination, troubleshooting can stop after final verification.

6. Compare With Known-Good Optical Components

When available, substitute approved known-good components one at a time, such as the light cable or compatible scope.

Avoid changing several components simultaneously because that makes the actual cause difficult to identify.

Expected outcome: The source of the illumination problem is isolated to a specific external component or remains associated with the system.

7. Verify Positioning and Field Illumination

On a suitable test target, confirm the scope and light delivery are positioned normally. Look for shadowing, partial obstruction, or apparent unevenness caused by scope orientation rather than electronic failure.

Expected outcome: Illumination is reasonably uniform for the test configuration. If repositioning or correcting an optical obstruction resolves the issue, troubleshooting can stop.

8. Verify AIM Selection and User-Accessible Configuration

Confirm the intended imaging mode is selected through normal clinical controls and that relevant user-accessible configuration has not been changed.

Do not enter unauthorized service menus or alter calibration values.

Expected outcome: AIM mode activates through the normal workflow and remains stable.

9. Perform Final Functional Verification

With the complete intended setup, verify:

Expected outcome: AIM functionality operates consistently. If all checks pass, the system can be returned to service according to facility policy and troubleshooting can stop.

If the Problem Persists

If compatible accessories, optical connections, light delivery, camera recognition, and normal configuration have been verified, the remaining cause may involve the light source, camera system, optical assembly, communication interface, calibration, or another service-level condition.

The affected system should be:

Do not attempt internal optical alignment or board-level repair. Complete appropriate functional verification before return to clinical use.

Knowing when to stop external troubleshooting is especially important when fluorescence guidance cannot be verified reliably.

Clinical Use Tip

Confirm the entire fluorescence imaging chain before the procedure rather than relying on successful standard white-light imaging alone.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"OR staff reported that AIM fluorescence mode activated but illumination was uneven across the surgical image."

Cause

What was observed during troubleshooting.

Example:
"Clinical Engineering found the issue followed the connected light cable while the camera and light source operated normally with a known-good cable."

Resolution

What action was taken.

Example:
"Replaced the affected light cable with an approved known-good cable and verified stable white-light and AIM illumination before return to service."

Helpful Details to Include (If Known)

Final Thought

Protect the clinical workflow first, then verify compatibility, connections, optics, positioning, and configuration before suspecting an internal fluorescence-system problem. Escalate when AIM performance cannot be verified and document exactly what was tested.

That is successful troubleshooting.