What This Guide Helps With
AIM fluorescence activation or illumination problems caused by setup, compatible accessories, connections, light delivery, contamination, positioning, or configuration.
Step-by-Step Troubleshooting
1. Protect the Patient and Maintain Visualization
If AIM fluorescence is required during an active procedure and is unavailable or unreliable, notify the clinical team. Do not repeatedly manipulate the system while fluorescence guidance is clinically required. Use a verified backup system or approved alternate workflow before troubleshooting.
Expected outcome: The clinical team has reliable visualization appropriate for the procedure.
2. Confirm the Exact AIM Failure
Determine whether AIM mode:
- Will not become available
- Will not activate
- Activates but produces weak fluorescence
- Produces uneven illumination
- Drops out intermittently
Document any displayed message and whether normal visible-light imaging remains functional.
Expected outcome: The failure is clearly defined and can be evaluated without patient dependence.
3. Verify the Complete Compatible Imaging Chain
Confirm that the connected camera head, camera control unit, light source, light cable, scope, and other required accessories are intended for the requested fluorescence workflow.
Do not assume AIM functionality simply because normal white-light video is present.
Expected outcome: The system is assembled with the appropriate compatible components. If correcting an accessory mismatch restores AIM operation, the issue is resolved and troubleshooting can stop after verification.
4. Check Camera and Light-Source Communication
Verify the camera head and light source are recognized normally. Inspect all external communication and video connections associated with the system.
Reseat accessible connections that can be safely disconnected and reconnect them fully.
Expected outcome: The camera system and light source remain connected and communicate without intermittent messages.
5. Inspect the Light Cable and Optical Connections
Check the light cable for:
- Loose attachment
- Damaged connectors
- Crushed or sharply bent sections
- Contamination on accessible optical surfaces
- Visible fiber damage when externally observable
Inspect the scope connection and other accessible optical interfaces according to approved cleaning practices.
Expected outcome: Light delivery is uniform and connections are secure. If cleaning or correctly seating an optical connection restores even illumination, troubleshooting can stop after final verification.
6. Compare With Known-Good Optical Components
When available, substitute approved known-good components one at a time, such as the light cable or compatible scope.
Avoid changing several components simultaneously because that makes the actual cause difficult to identify.
Expected outcome: The source of the illumination problem is isolated to a specific external component or remains associated with the system.
7. Verify Positioning and Field Illumination
On a suitable test target, confirm the scope and light delivery are positioned normally. Look for shadowing, partial obstruction, or apparent unevenness caused by scope orientation rather than electronic failure.
Expected outcome: Illumination is reasonably uniform for the test configuration. If repositioning or correcting an optical obstruction resolves the issue, troubleshooting can stop.
8. Verify AIM Selection and User-Accessible Configuration
Confirm the intended imaging mode is selected through normal clinical controls and that relevant user-accessible configuration has not been changed.
Do not enter unauthorized service menus or alter calibration values.
Expected outcome: AIM mode activates through the normal workflow and remains stable.
9. Perform Final Functional Verification
With the complete intended setup, verify:
- Normal visible-light imaging
- AIM availability
- AIM activation
- Stable illumination
- No unexpected mode dropout
- Stable camera and light-source communication
Expected outcome: AIM functionality operates consistently. If all checks pass, the system can be returned to service according to facility policy and troubleshooting can stop.
If the Problem Persists
If compatible accessories, optical connections, light delivery, camera recognition, and normal configuration have been verified, the remaining cause may involve the light source, camera system, optical assembly, communication interface, calibration, or another service-level condition.
The affected system should be:
- Removed from service when AIM functionality is required for intended clinical use
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate Stryker documentation and approved test equipment
- Repaired, calibrated, or configured only by qualified personnel
Do not attempt internal optical alignment or board-level repair. Complete appropriate functional verification before return to clinical use.
Knowing when to stop external troubleshooting is especially important when fluorescence guidance cannot be verified reliably.
Clinical Use Tip
Confirm the entire fluorescence imaging chain before the procedure rather than relying on successful standard white-light imaging alone.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported that AIM fluorescence mode activated but illumination was uneven across the surgical image."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering found the issue followed the connected light cable while the camera and light source operated normally with a known-good cable."
Resolution
What action was taken.
Example:
"Replaced the affected light cable with an approved known-good cable and verified stable white-light and AIM illumination before return to service."
Helpful Details to Include (If Known)
- Whether AIM was unavailable, weak, intermittent, or uneven
- Connected camera head and light source
- Light-cable condition
- Scope used during testing
- Optical connection condition
- Known-good substitution results
- Whether normal white-light imaging worked
- Any displayed message
- Final AIM verification result
Final Thought
Protect the clinical workflow first, then verify compatibility, connections, optics, positioning, and configuration before suspecting an internal fluorescence-system problem. Escalate when AIM performance cannot be verified and document exactly what was tested.
That is successful troubleshooting.