On this page
Asset Type
Manufacturer
Model
What This Guide Helps With
Video artifacts or interruptions caused by cables, connectors, routing equipment, displays, camera connections, signal conversion, or system instability.
Step-by-Step Troubleshooting
1. Protect the Patient and Establish Stable Visualization
If lines, sparkles, severe latency, or black frames occur during a procedure, notify the clinical team and transfer visualization to a verified stable video path or backup system.
Do not troubleshoot intermittent procedural video while the surgical team depends on it.
Expected outcome: Reliable visualization is established before technical testing.
2. Characterize the Artifact
Determine whether the problem consists of:
- Horizontal or vertical lines
- Sparkling pixels
- Delayed video
- Intermittent black frames
- Brief signal loss
- Image corruption during cable movement
Note whether the artifact appears on every display or only one destination.
Expected outcome: The video symptom and affected path are clearly documented.
3. Verify the Source Image
Connect the camera system to a known-good compatible display using the simplest available direct connection.
Expected outcome: A stable direct image indicates the camera source is functioning and shifts troubleshooting toward external routing or display infrastructure.
4. Inspect and Reseat Video Connections
Inspect all accessible video connectors for:
- Loose seating
- Physical damage
- Bent connector housings
- Cable tension
- Contamination
- Damaged strain reliefs
Reseat connections securely.
Expected outcome: The artifact disappears and does not return. If the image remains stable through verification, troubleshooting can stop.
5. Substitute the Video Cable
Use a known-good approved video cable in place of the suspect cable.
Expected outcome: If the artifact disappears, replace the defective cable and verify the complete path.
6. Bypass Switchers, Extenders, and Integration Hardware
Temporarily bypass intermediate equipment such as:
- Video switchers
- Wall interfaces
- Extenders
- Converters
- Recorders
- OR integration systems
Connect source to display directly when practical.
Expected outcome: A stable direct image isolates the fault to the external routing path or infrastructure.
7. Test Another Display
Connect the same verified source and cable to a second known-good compatible display.
Expected outcome: If the artifact follows the original display, refer the display for evaluation. If the artifact remains, continue isolating the source path.
8. Check Camera-Head and Camera-Cable Stability
Confirm the camera head remains recognized while the image artifact occurs. Gently reposition the camera cable during bench testing and observe whether video corruption coincides with camera communication changes.
Expected outcome: Camera recognition and video remain stable. If the failure follows camera-cable movement, remove the affected camera head from service.
9. Check for Environmental or Infrastructure Factors
Look for recently added adapters, unsupported converters, excessive cable stress, damaged wall interfaces, or equipment changes in the OR video chain.
Compare the setup with a known-working room or known-working path when practical.
Expected outcome: Infrastructure-related causes are identified or ruled out.
10. Perform Final Functional Verification
Verify the intended clinical video path for sufficient time to demonstrate:
- No lines or sparkles
- No intermittent black frames
- Normal responsiveness
- Stable camera recognition
- Stable routing through required integration equipment
Expected outcome: Video remains continuously stable. Troubleshooting can stop and the system may be returned to service.
If the Problem Persists
If direct output, cables, displays, routing equipment, camera connections, and external infrastructure have been checked, the remaining cause may involve internal video processing, camera-head electronics, output hardware, system software, or another service-level condition.
The affected equipment should be:
- Removed from service
- Labeled Out of Service
- Sent for repair or bench evaluation
- Evaluated using appropriate Stryker documentation and approved test equipment
- Repaired or configured only by qualified personnel
Coordinate with OR integration or audiovisual support if the failure has been isolated outside the camera platform. Verify the complete clinical video path after repair.
Knowing when intermittent video has not been conclusively corrected is proper troubleshooting.
Clinical Use Tip
An intermittent black frame that appears only occasionally still represents an unreliable visualization path and should not be dismissed because the image returns.
Work Order Documentation (CCR Method)
CCR = Complaint, Cause, Resolution
Complaint
What was reported by the clinical staff.
Example:
"OR staff reported intermittent sparkles and brief black frames on the primary monitor from the Stryker 1788 camera system."
Cause
What was observed during troubleshooting.
Example:
"Clinical Engineering reproduced the artifact through the installed external video cable while direct testing with a known-good cable remained stable."
Resolution
What action was taken.
Example:
"Replaced the defective video cable and verified continuous artifact-free video through the complete OR display path."
Helpful Details to Include (If Known)
- Type of video artifact
- Displays affected
- Whether direct output was stable
- Cable and connector condition
- Known-good cable result
- Known-good display result
- Switchers, extenders, or converters tested
- Camera recognition during the event
- Whether movement triggers the fault
- Final video verification result
Final Thought
Intermittent video requires the complete signal path to be tested systematically. Verify the source, then cables, routing equipment, and displays before assuming internal failure, and do not return a system to use until stability has been demonstrated.
That is successful troubleshooting.