Stryker 6090 / 6500 Series

Battery or Power Backup Failure

On this page

Asset Type

Stretcher

Manufacturer

Stryker

Model

6090 / 6500 Series

What This Guide Helps With

Stretcher battery not charging, no power to electric functions, or backup power failure during transport.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Expected: Patient care continues safely
Why it matters: Loss of powered functions (hi-lo, controls) can impact safe transport and positioning

Verify the Reported Issue

Expected: Clear confirmation of battery-related failure
Why it matters: Distinguishes between total power failure and battery backup issues

Check AC Power Connection (If Equipped)

Expected: Indicator lights or system powers on
Why it matters: Rules out external power supply issues before suspecting battery failure

Inspect Battery Charging Indicators

Expected: Charging indicator responds when plugged in
Why it matters: Helps determine if the charger circuit is working

Check for Blown Fuses or Resettable Breakers (External Only)

Expected: Power restored if protection device was tripped
Why it matters: Overcurrent protection can disable charging and mimic battery failure

Perform a Power Cycle

Expected: System may reset and restore function
Why it matters: Clears minor control or charging faults

Evaluate Battery Condition

Expected: Battery should hold charge and power basic functions
Why it matters: Batteries degrade over time and are a common failure point

Check for Physical Signs of Battery Issues

Expected: No abnormal signs
Why it matters: Indicates unsafe battery condition requiring immediate removal from service

If the Problem Persists

If all external power sources, connections, and reset attempts have been verified, the issue is likely internal (battery pack, charger board, or wiring).

Remove the stretcher from service, label it Out of Service, and send for repair or bench evaluation.

Knowing when to stop prevents unnecessary risk and ensures proper repair handling.

Clinical Use Tip

Do not rely on a stretcher with known battery issues during patient transport. Always verify functionality before moving a patient, especially for powered positioning. Move the patient to a backup device first to ensure therapy continuity.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Stretcher will not power on or loses power when unplugged."

Cause

What was observed during troubleshooting.

Example:
"Failed internal battery not holding charge."

Resolution

What action was taken.

Example:
"Verified AC power, inspected connections, removed stretcher from service for battery replacement."

Helpful Details to Include (If Known)

Final Thought

Battery and power issues are often straightforward when approached logically. Always rule out external power and connection problems first, then identify battery failure as a common cause. Prioritize patient safety and escalate appropriately when internal repair is required. Clear documentation ensures continuity and accountability.

That is successful troubleshooting.