Stryker 6090 / 6500 Series

Brake or Caster Locking Failure

On this page

Asset Type

Stretcher

Manufacturer

Stryker

Model

6090 / 6500 Series

What This Guide Helps With

Brake pedals not engaging, stretcher rolling when brakes applied, or caster locking system not functioning properly.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Why it matters: Uncontrolled movement can lead to patient falls or staff injury.

Verify the Reported Issue

Expected: All casters should firmly lock when brakes are engaged.

Why it matters: Confirms whether it’s a full system failure or isolated to one wheel.

Check Brake Pedal Linkage and Movement

Expected: Pedal moves freely and fully engages/disengages brakes.

Why it matters: Mechanical linkage issues are a common external failure point.

Inspect for Obstructions or Debris in Casters

Expected: Casters rotate and lock freely without obstruction.

Why it matters: Debris can prevent full brake engagement.

Check Individual Casters for Locking Function

Expected: All casters should resist movement when locked.

Why it matters: A single failed caster can make the entire system appear ineffective.

Inspect Brake Rods and Linkages Under the Frame

Expected: Linkage is intact and moves when pedal is pressed.

Why it matters: Mechanical disconnection prevents brake force from reaching casters.

Check for Uneven Floor Contact or Load Distribution

Expected: Even contact allows all brakes to engage properly.

Why it matters: Uneven load can prevent certain casters from locking.

Test Brake Function Under Load

Expected: Stretcher should not roll under normal force.

Why it matters: Confirms real-world brake performance, not just unloaded function.

If the Problem Persists

If all external checks and adjustments have been completed and the brake system still fails, the issue is likely internal (worn brake assemblies, internal linkage failure, or caster mechanism failure).

Remove the stretcher from service, label it Out of Service, and send it for repair or bench evaluation.

Knowing when to stop is proper troubleshooting.

Clinical Use Tip

Never troubleshoot brake issues with a patient on the stretcher. Always transfer the patient first to prevent unintended movement.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Staff reported stretcher brakes not holding and unit rolling when parked."

Cause

What was observed during troubleshooting.

Example:
"Debris buildup and one caster not locking properly due to worn brake mechanism."

Resolution

What action was taken.

Example:
"Removed debris, identified faulty caster, and removed stretcher from service for caster replacement."

Helpful Details to Include (If Known)

Final Thought

Brake failures are high-risk issues that require immediate attention. Always start with simple mechanical checks, but do not hesitate to escalate when braking reliability is compromised. Proper documentation and timely removal from service protect both patients and staff.

That is successful troubleshooting.