Stryker 6090 / 6500 Series

Trendelenburg or Backrest Not Adjusting

On this page

Asset Type

Stretcher

Manufacturer

Stryker

Model

6090 / 6500 Series

What This Guide Helps With

Stretcher backrest or Trendelenburg function not moving, stuck, or slow due to mechanical, hydraulic, or control-related external issues.

Step-by-Step Troubleshooting

Ensure Patient Safety First

Why it matters: Positioning functions are critical for patient care and safety.

Verify the Reported Issue

Expected: Identify if issue is isolated or system-wide.

Why it matters: Helps determine mechanical vs hydraulic vs control issue.

Check for Mechanical Obstruction

Expected: No obstruction present.

Why it matters: Even small obstructions can prevent movement.

Verify Control Lever/Pedal Operation

Expected: Lever/pedal moves freely and returns to neutral.

Why it matters: Mechanical linkage failure is a common cause.

Check Hydraulic System Response

Expected: Smooth movement when control is engaged.

Why it matters: No response may indicate hydraulic failure or low fluid.

Inspect for Hydraulic Leaks

Expected: No visible leaks.

Why it matters: Loss of hydraulic fluid prevents proper positioning.

Check Load and Weight Distribution

Expected: Movement improves if overload was the issue.

Why it matters: Excess weight can prevent hydraulic actuation.

Cycle the Function Multiple Times

Expected: Movement may restore if system was stiff or not used frequently.

Why it matters: Helps free minor internal sticking.

Compare with a Known Working Unit (If Available)

Expected: Identify abnormal resistance or behavior.

Why it matters: Confirms whether issue is truly abnormal.

If the Problem Persists

If all external checks, controls, and obstructions have been ruled out, the issue is likely internal to the hydraulic system or mechanical linkage.

Knowing when to stop prevents unnecessary risk and reflects proper troubleshooting.

Clinical Use Tip

Never attempt to troubleshoot positioning functions while a patient is on the stretcher. Always transfer the patient first to avoid injury.

Work Order Documentation (CCR Method)

CCR = Complaint, Cause, Resolution

Complaint

What was reported by the clinical staff.

Example:
"Stretcher backrest and Trendelenburg functions not adjusting per clinical staff report."

Cause

What was observed during troubleshooting.

Example:
"Hydraulic system not responding; no external obstruction found during inspection."

Resolution

What action was taken.

Example:
"Removed from service and sent for hydraulic system repair and evaluation."

Helpful Details to Include (If Known)

Final Thought

Positioning failures on stretchers should always be approached methodically—starting with simple mechanical checks before assuming hydraulic failure. Patient safety and proper escalation are key, and clear documentation ensures efficient repair follow-up.

That is successful troubleshooting.